§792.63 Maintenance and calibration of equipment.
(a) Equipment shall be adequately inspected, cleaned, and maintained. Equipment used for the generation, measurement, or assessment of data shall be adequately tested, calibrated, and/or standardized.
(b) The written standard operating procedures required under §792.81(b)(11) shall set forth in sufficient detail the methods, materials, and schedules to be used in the routine inspection, cleaning, maintenance, testing, calibration, and/or standardization of equipment, and shall specify, when appropriate, remedial action to be taken in the event of failure or malfunction of equipment. The written standard operating procedures shall designate the person responsible for the performance of each operation.
(c) Written records shall be maintained of all inspection, maintenance,
testing, calibrating, and/or standardizing operations. These records, containing
the date of the operation, shall describe whether the maintenance operations
were routine and followed the written standard operating procedures. Written
records shall be kept of nonroutine repairs performed on equipment as a result
of failure and malfunction. Such records shall document the nature of the
defect, how and when the defect was discovered, and any remedial action taken in
response to the defect.