|
||||||
|---|---|---|---|---|---|---|
| Home | 29 CFR | 40 CFR | 49 CFR | Federal Register | MSDS Search | Purchase CFR e-Books |
§53.62 Test procedure: Full wind tunnel test.
(a) Overview. The full wind tunnel test evaluates the effectiveness of
the candidate sampler at 2 km/hr and 24 km/hr for aerosols of the size specified
in table F-2 of this subpart (under the heading, "Full Wind Tunnel Test"). For
each wind speed, a smooth curve is fit to the effectiveness data and corrected
for the presence of multiplets in the wind tunnel calibration aerosol. The
cutpoint diameter (D (b) Technical definitions. Effectiveness is the ratio (expressed as a
percentage) of the mass concentration of particles of a specific size reaching
the sampler filter or filters to the mass concentration of particles of the same
size approaching the sampler.
(c) Facilities and equipment required -- (1) Wind tunnel. The
particle delivery system shall consist of a blower system and a wind tunnel
having a test section of sufficiently large cross-sectional area such that the
test sampler, or portion thereof, as installed in the test section for testing,
blocks no more than 15 percent of the test section area. The wind tunnel blower
system must be capable of maintaining uniform wind speeds at the 2 km/hr and 24
km/hr in the test section.
(2) Aerosol generation system. A vibrating orifice aerosol generator
shall be used to produce monodisperse solid particles of ammonium fluorescein
with equivalent aerodynamic diameters as specified in table F-2 of this subpart.
The geometric standard deviation for each particle size generated shall not
exceed 1.1 (for primary particles) and the proportion of multiplets (doublets
and triplets) in all test particle atmosphere shall not exceed 10 percent of the
particle population. The aerodynamic particle diameter, as established by the
operating parameters of the vibrating orifice aerosol generator, shall be within
the tolerance specified in table F-2 of this subpart.
(3) Particle size verification equipment. The size of the test
particles shall be verified during this test by use of a suitable instrument
(e.g., scanning electron microscope, optical particle sizer, time-of-flight
apparatus). The instrument must be capable of measuring solid and liquid test
particles with a size resolution of 0.1 μm or less. The accuracy of the particle
size verification technique shall be 0.15 μm or better.
(4) Wind speed measurement. The wind speed in the wind tunnel shall be
determined during the tests using an appropriate technique capable of a
precision of 2 percent and an accuracy of 5 percent or better (e.g., hot-wire
anemometry). For the wind speeds specified in table F-2 of this subpart, the
wind speed shall be measured at a minimum of 12 test points in a cross-sectional
area of the test section of the wind tunnel. The mean wind speed in the test
section must be within ±10 percent of the value specified in table F-2 of this
subpart, and the variation at any test point in the test section may not exceed
10 percent of the measured mean.
(5) Aerosol rake. The cross-sectional uniformity of the particle
concentration in the sampling zone of the test section shall be established
during the tests using an array of isokinetic samplers, referred to as a rake.
Not less than five evenly spaced isokinetic samplers shall be used to determine
the particle concentration spatial uniformity in the sampling zone. The sampling
zone shall be a rectangular area having a horizontal dimension not less than 1.2
times the width of the test sampler at its inlet opening and a vertical
dimension not less than 25 centimeters.
(6) Total aerosol isokinetic sampler. After cross-sectional uniformity
has been confirmed, a single isokinetic sampler may be used in place of the
array of isokinetic samplers for the determination of particle mass
concentration used in the calculation of sampling effectiveness of the test
sampler in paragraph (d)(5) of this section. In this case, the array of
isokinetic samplers must be used to demonstrate particle concentration
uniformity prior to the replicate measurements of sampling effectiveness.
(7) Fluorometer. A fluorometer used for quantifying extracted aerosol
mass deposits shall be set up, maintained, and calibrated according to the
manufacturer's instructions. A series of calibration standards shall be prepared
to encompass the minimum and maximum concentrations measured during
size-selective tests. Prior to each calibration and measurement, the fluorometer
shall be zeroed using an aliquot of the same solvent used for extracting aerosol
mass deposits.
(8) Sampler flow rate measurements. All flow rate measurements used to
calculate the test atmosphere concentrations and the test results must be
accurate to within ±2 percent, referenced to a NIST-traceable primary standard.
Any necessary flow rate measurement corrections shall be clearly documented. All
flow rate measurements shall be performed and reported in actual volumetric
units.
(d) Test procedures -- (1) Establish and verify wind speed. (i)
Establish a wind speed specified in table F-2 of this subpart.
(ii) Measure the wind speed at a minimum of 12 test points in a
cross-sectional area of the test section of the wind tunnel using a device as
described in paragraph (c)(4) of this section.
(iii) Verify that the mean wind speed in the test section of the wind tunnel
during the tests is within 10 percent of the value specified in table F-2 of
this subpart. The wind speed measured at any test point in the test section
shall not differ by more than 10 percent from the mean wind speed in the test
section.
(2) Generate aerosol. (i) Generate particles of a size specified in
table F-2 of this subpart using a vibrating orifice aerosol generator.
(ii) Check for the presence of satellites and adjust the generator as
necessary.
(iii) Calculate the physical particle size using the operating parameters of
the vibrating orifice aerosol generator and record.
(iv) Determine the particle's aerodynamic diameter from the calculated
physical diameter and the known density of the generated particle. The
calculated aerodynamic diameter must be within the tolerance specified in table
F-2 of this subpart.
(3) Introduce particles into the wind tunnel. Introduce the generated
particles into the wind tunnel and allow the particle concentration to
stabilize.
(4) Verify the quality of the test aerosol. (i) Extract a
representative sample of the aerosol from the sampling test zone and measure the
size distribution of the collected particles using an appropriate sizing
technique. If the measurement technique does not provide a direct measure of
aerodynamic diameter, the geometric mean aerodynamic diameter of the challenge
aerosol must be calculated using the known density of the particle and the
measured mean physical diameter. The determined geometric mean aerodynamic
diameter of the test aerosol must be within 0.15 μm of the aerodynamic diameter
calculated from the operating parameters of the vibrating orifice aerosol
generator. The geometric standard deviation of the primary particles must not
exceed 1.1.
(ii) Determine the population of multiplets in the collected sample. The
multiplet population of the particle test atmosphere must not exceed 10 percent
of the total particle population.
(5) Aerosol uniformity and concentration measurement. (i) Install an
array of five or more evenly spaced isokinetic samplers in the sampling zone
(paragraph (c)(5) of this section). Collect particles on appropriate filters
over a time period such that the relative error of the measured particle
concentration is less than 5.0 percent.
(ii) Determine the quantity of material collected with each isokinetic
sampler in the array using a calibrated fluorometer. Calculate and record the
mass concentration for each isokinetic sampler as:
where: i = replicate number;
j = isokinetic sampler number;
M Q = isokinetic sampler volumetric flow rate; and
t = sampling time. (iii) Calculate and record the mean mass concentration as:
where: i = replicate number;
j = isokinetic sampler number; and
n = total number of isokinetic samplers. (iv) Precision calculation. (A) Calculate the coefficient of variation of the
mass concentration measurements as:
where: i = replicate number;
j = isokinetic sampler number; and
n = total number of isokinetic samplers. (B) If the value of CV (6) Alternative measure of wind tunnel total concentration. If a
single isokinetic sampler is used to determine the mean aerosol concentration in
the wind tunnel, install the sampler in the wind tunnel with the sampler nozzle
centered in the sampling zone (paragraph (c)(6) of this section).
(i) Collect particles on an appropriate filter over a time period such that
the relative error of the measured concentration is less than 5.0 percent.
(ii) Determine the quantity of material collected with the isokinetic sampler
using a calibrated fluorometer.
(iii) Calculate and record the mass concentration as C (iv) Remove the isokinetic sampler from the wind tunnel.
(7) Measure the aerosol with the candidate sampler. (i) Install the
test sampler (or portion thereof) in the wind tunnel with the sampler inlet
opening centered in the sampling zone. To meet the maximum blockage limit of
paragraph (c)(1) of this section or for convenience, part of the test sampler
may be positioned external to the wind tunnel provided that neither the geometry
of the sampler nor the length of any connecting tube or pipe is altered. Collect
particles for a time period such that the relative error of the measured
concentration is less than 5.0 percent.
(ii) Remove the test sampler from the wind tunnel.
(iii) Determine the quantity of material collected with the test sampler
using a calibrated fluorometer. Calculate and record the mass concentration for
each replicate as:
where: i = replicate number;
M Q = candidate sampler volumetric flow rate; and
t = sampling time. (iv)(A) Calculate and record the sampling effectiveness of the candidate
sampler as:
where: i = replicate number. (B) If a single isokinetic sampler is used for the determination of particle
mass concentration, replace C (8) Replicate measurements and calculation of mean sampling
effectiveness. (i) Repeat steps in paragraphs (d)(5) through (d)(7) of this
section, as appropriate, to obtain a minimum of three valid replicate
measurements of sampling effectiveness.
(ii) Calculate and record the average sampling effectiveness of the test
sampler for the particle size as:
where: i = replicate number; and
n = number of replicates. (iii) Sampling effectiveness precision. (A) Calculate and record the
coefficient of variation for the replicate sampling effectiveness measurements
of the test sampler as:
where: i = replicate number, and
n = number of replicates. (B) If the value of CV (9) Repeat steps in paragraphs (d)(2) through (d)(8) of this section until
the sampling effectiveness has been measured for all particle sizes specified in
table F-2 of this subpart.
(10) Repeat steps in paragraphs (d)(1) through (d)(9) of this section until
tests have been successfully conducted for both wind speeds of 2 km/hr and 24
km/hr.
(e) Calculations -- (1) Graphical treatment of effectiveness
data. For each wind speed given in table F-2 of this subpart, plot the
particle average sampling effectiveness of the candidate sampler as a function
of aerodynamic particle diameter (D (2) Cutpoint determination. For each wind speed determine the sampler
Dp (3) Expected mass concentration calculation. For each wind speed,
calculate the estimated mass concentration measurement for the test sampler
under each particle size distribution (Tables F-4, F-5, and F-6 of this subpart)
and compare it to the mass concentration predicted for the reference sampler as
follows:
(i) Determine the value of corrected effectiveness using the best-fit,
multiplet-corrected curve at each of the particle sizes specified in the first
column of table F-4 of this subpart. Record each corrected effectiveness value
as a decimal between 0 and 1 in column 2 of table F-4 of this subpart.
(ii) Calculate the interval estimated mass concentration measurement by
multiplying the values of corrected effectiveness in column 2 by the interval
mass concentration values in column 3 and enter the products in column 4 of
table F-4 of this subpart.
(iii) Calculate the estimated mass concentration measurement by summing the
values in column 4 and entering the total as the estimated mass concentration
measurement for the test sampler at the bottom of column 4 of table F-4 of this
subpart.
(iv) Calculate the estimated mass concentration ratio between the candidate
method and the reference method as:
where: C C (v) Repeat steps in paragraphs (e) (1) through (e)(3) of this section for
tables F-5 and F-6 of this subpart.
(f) Evaluation of test results. The candidate method passes the wind
tunnel effectiveness test if the R