 |
BENZENE |
 |
| Method no.: |
1005 |
|
| Control no.: |
T-1005-FV-01-0111-M |
|
| Target concentration: |
1 ppm |
|
| OSHA PEL: |
1 ppm (3.19 mg/m3)(TWA); 5 ppm (16.0
mg/m3)(15-min STEL); 50 ppm (160 mg/m3)(10-min
Peak) in General Industry (29 CFR 1910.1028) 10 ppm (31.9
mg/m3)(TWA); 25 ppm (79.8 mg/m3)(15-min
Ceiling) in sectors excluded from General Industry (29 CFR
1926.1028) 1 ppm (3.19 mg/m3)(TWA); 5 ppm (16.0
mg/m3)(15-min STEL); 50 ppm (160 mg/m3)(10-min
Peak) in Shipyard and Construction Industries |
|
| ACGIH TLV: |
0.5 ppm (1.6 mg/m3) TWA; 2.5 ppm (8.0
mg/m3)(15-min STEL) (Skin) |
|
| Procedure: |
Active samples are collected by drawing workplace air
through charcoal tubes with personal sampling pumps. Diffusive
samples are collected by exposing either SKC 575-002 Passive
Samplers or 3M 3520 Organic Vapor Monitors (OVM) to workplace air.
Samples are extracted with carbon disulfide and analyzed by GC using
a flame ionization detector (FID). |
|
Recommended sampling time and sampling
rate: Charcoal tubes: |
240 min at 50 mL/min (12 L)(TWA); 10 min at 50 mL/min (0.5
L)(Peak); 15 min at 50 mL/min (0.75 L)(STEL) |
|
| SKC 575-002 Passive Sampler and 3M 3520 OVM: |
240 min (TWA); 10 min (Peak); 15 min (STEL)
(Note: SKC 575-002 Passive Samplers and 3M 3520 OVMs must be exposed
for at least 10 min) |
|
| Reliable quantitation limit (RQL) and standard error
of estimate (SEE): |
|
| Media |
RQL |
SEE |
|
(ppb) |
(µg/m³) |
(%) |
|
| Charcoal tube |
3.32 |
11 |
5.1 |
| SKC 575-002 Passive Sampler |
4.25 |
14 |
8.8 |
| 3M 3520 OVM |
3.55 |
11 |
7.4 |
| |
| *For samples where sampling site atmospheric pressure
and temperature are known. When either or both of these values are
unknown, see Section 4.4 for applicable standard errors of
estimate. |
|
| Special requirements: |
Report sampling site atmospheric pressure and temperature when
using diffusive samplers |
|
| Status of method: |
Evaluated method. This method has been subjected to the
established evaluation procedures of the Methods Development
Team. |
|
| November 2001 |
Mary E. Eide |
Methods Development Team Industrial Hygiene Chemistry
Division OSHA Salt Lake Technical Center Salt Lake City UT
84115-1802
-
General Discussion
1.1 Background
1.1.1 History
The collection of benzene using charcoal tubes (SKC lot 107) was
evaluated in OSHA Method 12.1 Since that time, many
diffusive samplers have come on the market, and are becoming more
popular for workplace sampling. The Methods Development Team at OSHA
Salt Lake Technical Center is in the process of validating passive
samplers for the top 10 organic chemicals. Benzene is in the top 10
organic chemicals requested, therefore an evaluation of its sampling
performance with passive samplers was performed. This method includes
two diffusive samplers, SKC 575-002 Passive Samplers and 3M 3520 OVMs,
along with charcoal tubes (SKC lot 2000). Benzene PELs include a TWA,
STEL, Ceiling, and Peak; therefore, these media were evaluated for
both short term and long term sampling, for the various PELs.
The tests for the determination of the sampling rate show that SKC
575-002 Passive Samplers and 3M 3520 OVMs had a faster sampling rate
in the first five minutes of sampling, than the rest of the times
sampled. The sampling rate for the subsequent time periods each
decreased until at 30 minutes it leveled off to the sampling rate
used. The sampling rate was significantly higher in the first 5
minutes of sampling than the rest of the sampling period, so it is
recommended that the sampling time for these passive samplers be at
least 10 minutes. The Peak PEL for benzene is defined as a 10-minute
sample.
Each of the media had good extraction efficiency and storage
stability. The use of a capillary column in the analysis allowed for
lower reliable quantitation limits in this method when compared to the
limits obtained with the packed column in OSHA Method 12.2
1.1.2 Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy.)
Benzene in high concentrations has narcotic effects similar to
toluene and other aromatics. Benzene is a myclotoxicant known to
supress bone marrow cell proliferation and to induce hematologic
disorders in humans and animals. Chronic exposure to benzene leads to
aplastic anemia, and may lead to leukemia after 6 months to 6 years of
chronic exposure. Benzene exposure can cause chromosomal aberrations
in animals and humans. Chronic benzene exposure has also been
associated with lung cancer in epidemiological studies. Benzene is
classified as a human carcinogen by American Conference of
Governmental Industrial Hygienists (ACGIH) and International Agency
for Research on Cancer (IARC).3 Benzene is classified as a
suspected human carcinogen by OSHA.4 A risk assessment for
benzene exposure performed by Rinsky et al., reported that a worker
exposed to 10 ppm benzene for 40 years was 155 times more likely to
die from leukemia than an unexposed worker. A worker exposed to 1 ppm
benzene was 1.7 times more likely to die from leukemia than an
unexposed worker.5 IARC has published a risk assesment
showing that workers exposed to a 10 ppm chronic exposure, had an
increase of 14-140 leukemia cases per 1000 people above the rate for
an unexposed worker. For 1 ppm chronic exposure, there was an increase
of 1.4-14 cases per 1000 above the rate for an unexposed
worker.6 The skin notation cited in the TLV is based on the
skin absorption rate of 0.05% when neat benzene is applied to the
skin, indicating benzene exposures through skin absorption can be
significant.7
1.1.3 Workplace exposure
Benzene is used in the manufacture of industrial chemicals, as a
solvent for waxes, resins, oils, natural rubber. Benzene may be
present in gasoline and other petroleum products up to 1%, as a
natural part of the cracking process.8 Benzene was the
16th highest chemical in the ranking of production by
volume for 1995.9 Benzene production was 2.412 billion
gallons in 2000.10
1.1.4 Physical properties and descriptive
information11,12,13
CAS number: IMIS
number14: molecular weight: boiling
point: melting point: specific gravity:
|
71-43-2 0320 78.11 80.1°C 5.5°C 0.879
at 20°C
|
vapor pressure: lower explosive
limit: flash point: odor:
appearance: molecular
formula: |
10 kPa at 20°C 1.5 to 8%
by volume 11°C (52°F) (cc) characteristic
aromatic colorless liquid C6H6
|
| synonyms: |
benzol; coal naphtha;
cyclohexatriene; phenyl hydride |
| solubility: |
slightly soluble in water;
miscible in alcohol, acetone. carbon disulfide, carbon
tetrachloride, chloroform, ether, glacial acetic acid, and
oils |
| structure: |
 |
|
| This method was evaluated according to the OSHA
SLTC "Evaluation Guidelines for Air Sampling Methods Utilizing
Chromatographic Analysis"15. The Guidelines define
analytical parameters, specify required laboratory tests,
statistical calculations and acceptance criteria. The analyte air
concentrations throughout this method are based on the recommended
sampling and analytical parameters. Air concentrations listed in ppm
are referenced to 25°C and 101.3 kPa (760 mmHg). |
|
1.2 Limit defining parameters
1.2.1 Detection limit of the analytical procedure
The detection limit of the analytical procedure is 3.12 pg. This is
the amount of analyte that will give a detector response that is
significantly different from the response of a reagent blank. (Section
4.1)
1.2.2 Detection limit of the overall procedure The detection
limits of the overall procedure are 38 ng/sample for charcoal tubes,
49 ng/sample for SKC 575-002 Passive Samplers, and 41 ng/sample for 3M
3520 OVMs. These are the amounts of benzene spiked on the respective
sampler that will give detector responses that are significantly
different from the responses of respective sampler blanks. (Section
4.2)
Table 1.2.2 Detection Limits of the Overall
Procedure
|
| sampler |
ng |
ppb |
µg/m3 |
|
charcoal tube SKC
575-002 3M 3520 |
38 49 41 |
0.99 1.28 1.07 |
3.17 4.08 3.42 |
|
1.2.3 Reliable quantitation limit
The reliable quantitation limits are 127 ng per sample (3.3 ppb or
11 g/m3) for charcoal tubes, 163 ng per sample (4.3 ppb or
14 g/m3) for SKC 575-002 Passive Samplers, and 136 ng per
sample (3.6 ppb or 11 µg/m3) for 3M 3520 OVMs. These are
the amounts of benzene spiked on the respective samplers that will
give detector responses that are considered the lower limits for
precise quantitative measurements. (Section 4.2)
Table 1.2.3 Reliable Quantitation Limits
|
| sampler |
ng |
ppb |
µg/m3 |
EE |
|
charcoal tube SKC
575-002 3M 3520 |
127 163 136 |
3.32
4.26 3.55 |
10.6
13.6 11.3 |
96.6
93.1 97.9 |
|
1.2.4 Instrument calibration
The standard error of estimate is 0.34 µg over the range of 9.9 to
79.1 µg. This range corresponds to 0.25 to 2 times the TWA target
concentration. (Section 4.3)
1.2.5 Precision
Charcoal Tubes
The precision of the overall procedure at the 95% confidence level
for the ambient temperature 19-day storage test for samples collected
from a dynamically generated atmosphere of 1.2 ppm (3.84
mg/m3) collected on charcoal tubes is ±9.94%. This includes
an additional 5% for sampling pump variability. (Section 4.4)
SKC 575-002 Passive Samplers
The precisions of the overall procedure at the 95% confidence level
for the ambient temperature 19-day storage test for samples collected
from a dynamically generated atmosphere of 1.2 ppm (3.84
mg/m3) collected on SKC 575-002 Passive Samplers are given
in Table 1.2.5.1. They each include an additional 8.7% for sampling
rate variability. There are different values given, depending on
whether both, either, or neither temperature (T) or
atmospheric pressure (P) are known at the sampling site. If
the sampling site temperature is unknown, it is assumed to be 22.2 ±
15°C (72 ± 27°F) and a variability of ±7.7% is included. If the
atmospheric pressure is not known, it is estimated from the sampling
site elevation and a variability of ±3% is included. (Section 4.4)
Table 1.2.5.1 Precision of the Overall
Procedure of SKC 575-002 Passive Samplers
|
| know conditions |
precision (±%) |
|
both T &
P only T only P neither T
nor P |
17.2 18.1 22.8 23.6 |
|
3M 3520 OVMs
The precisions of the overall procedure at the 95% confidence level
for the ambient temperature 19-day storage test for samples collected
from a dynamically generated atmosphere of 1.2 ppm (3.84
mg/m3) collected on 3M 3520 OVMs are given in Table
1.2.5.2. They each include an additional 6.4% for sampling rate
variability. There are different values given, depending on whether
both, either, or neither temperature (T) or atmospheric
pressure (P) are known at the sampling site. If the sampling
site temperature is unknown, it is assumed to be 22.2 ± 15°C (72 ±
27°F) and a variability of ±7.7% is included. If the atmospheric
pressure is not known, it is estimated from the sampling site
elevation and a variability of ±3% is included. (Section 4.4)
Table 1.2.5.2 Precision of Overall Procedure for 3M 3520
OVMs
|
| know conditions |
precision (±%) |
|
both T &
P only T only P neither T
nor P |
12.6 13.9 19.6 20.6 |
|
1.2.6 Recovery
The recovery of benzene from samples used in a 19-day storage test
remained above 98.0%, 96.5%, and 96.2% when the samples were stored at
23°C for charcoal tubes, SKC 575-002 Passive Samplers, and 3M 3520
OVMs, respectively. (Section 4.5)
1.2.7 Reproducibility
Six samples for each of the three types of samplers were collected
from a controlled test atmosphere and submitted for analysis by the
OSHA Salt Lake Technical Center. The samples were analyzed according
to a draft copy of this procedure after 7 days of storage at 4°C. No
individual sample result deviated from its theoretical value by more
than the precision reported in Section 1.2.5. (Section
4.6)
- Sampling Procedure
All safety practices that apply to the work area being sampled should
be followed. The sampling equipment should be attached to the worker in
a manner that will not interfere with work performance or
safety.
2.1 Apparatus
2.1.1 Charcoal tubes
Samples are collected with 7-cm × 4-mm i.d. × 6-mm o.d. glass
sampling tubes packed with two sections of coconut shell charcoal. The
front section contains 100 mg and the back section contains 50 mg of
charcoal. The sections are held in place with glass wool and
polyurethane plugs. For this evaluation, commercially prepared
charcoal tubes were purchased from SKC, Inc. (catalog no. 226-01,
Anasorb CSC, lot 2000).
Samples are collected using a personal sampling pump calibrated,
with the sampling device attached, to within ±5% of the recommended
flow rate.
2.1.2 SKC 575-002 Passive Samplers and 3M 3520 OVMs
Samples are collected with either SKC 575-002 Passive Samplers, or
with 3M 3520 OVMs. Samplers were purchased from SKC, Inc. (catalog no.
575-002, contains 500 mg of Anasorb 747) or from 3M (catalog no. 3520,
contains two charcoal adsorbent pads).
A thermometer and barometer to determine the sampling site air
temperature and atmospheric pressure.
2.2 Reagents
None required
2.3 Technique
2.3.1 Charcoal tubes
Immediately before sampling, break off the ends of the flame-sealed
tube to provide an opening approximately half the internal diameter of
the tube at each end. Wear eye protection when breaking ends. Use tube
holders to minimize the hazard of broken glass. All tubes should be
from the same lot.
The smaller section of the adsorbent tube is used as a back-up and
is positioned nearest the sampling pump. Attach the tube holder to the
sampling pump so that the adsorbent tube is in an approximately
vertical position with the inlet facing down during sampling. Position
the sampling pump, tube holder and tubing so they do not impede work
performance or safety. Use tube holder to minimize the hazard to the
worker from the broken end of the tube.
Draw the air to be sampled directly into the inlet of the tube
holder. The air being sampled should not be passed through any hose or
tubing before entering the sampling tube.
After sampling for the appropriate time, remove the adsorbent tube
and seal it with plastic end caps. Seal each sample end-to-end with an
OSHA 21 form as soon as possible.
Submit at least one blank sample with each set of samples. Handle
the blank sampler in the same manner as the other samples except draw
no air through it.
Record sample air volume (liters), sampling time (minutes) and
sampling rate (mL/min) for each sample, along with any potential
interferences on the OSHA 91A form.
Submit the samples to the laboratory for analysis as soon as
possible after sampling. If delay is unavoidable, store the samples in
a refrigerator. Ship any bulk samples separate from the air
samples.
2.3.2 SKC 575-002 Passive Samplers (In general, follow the
manufacturer's instructions.)
Remove the sampler enclosed in an air-tight clear bag from the
container, just before sampling is to begin. Caution- The
sampler begins to sample immediately after the clear plastic bag is
opened. Keep the O-ring, press-on cover, cover retainer, port
plugs and PTFE tube for later use.
Record the start time on the sampler label or on the OSHA 91A
form.
Attach the sampler to the worker near his/her breathing zone with
the perforations in the sampler facing forward. Assure that the area
directly in front of the sampler is unobstructed throughout the
sampling period.
At the end of the sampling period, immediately detach the sampler
from the worker and attach the cover with the O-ring in place onto the
sampler using the cover retainer. Visually inspect the O-ring to be
sure it is forming a proper seal around the entire circumference of
the sampler. Record the stop time on sampler label or on OSHA 91A
form.
Prepare a blank by removing an unused sampler from its clear
package and immediately attaching a cover with the O-ring in place
onto it.
Seal each sampler with an OSHA 21 form.
Verify that the sampling times are properly recorded on the OSHA
91A form for each sample. Also, identify blank samples on this
form.
Record the room temperature and atmospheric pressure of the
sampling site on the OSHA 91A form.
List any compounds that could be considered potential
interferences, especially solvents, that are being used in the
sampling area.
Submit the samples to the laboratory for analysis as soon as
possible after sampling. If delay is unavoidable, store the samples in
a refrigerator. Ship any bulk samples separate from the air samples.
Include all port plugs and PTFE tubes which will be used in the
laboratory analyses. Ship any bulk sample(s) in a container separate
from the air samples.
2.3.3 3M OVMs (In general, follow the manufacture's instructions
supplied with the samplers.)
The samplers come individually sealed in small metal cans. When
ready to begin sampling, remove the plastic lid from the can and lift
up on the revealed ring. Pull back on the ring to open the can.
Discard the metal top of the can and remove the sampler.
Caution - The sampler begins to sample immediately after the
can is unsealed.
Keep the two closure caps with attached port plugs, cup and PTFE
tubes in the can for later use. Close the can with the plastic
lid.
Record the start time on the back of the sampler or on the OSHA 91A
form.
Attach the sampler to the worker near his/her breathing zone with
the white face forward. Assure that the area directly in front of the
sampler is unobstructed throughout the sampling period. Do not remove
the white film and ring from the sampler until the sampling period is
terminated.
At the end of the sampling period, detach the sampler from the
worker and remove the white film and retaining ring. Immediately snap
a closure cap onto the primary (top) section of the sampler (where the
white film and ring were removed). It is critical that this step be
done as quickly as possible because the sampling rate is more than
five times faster without the white film in place, which can be an
important consideration, especially for short-term sampling. Assure
that the attached port plugs are placed firmly into the port holes.
The white film and ring can be discarded. Record the stop time on the
back of the sampler and on the OSHA 91A form.
The following steps should be performed in a low background area
for a set of samplers as soon as possible after sampling.
Ready a blank by removing the white film and ring and attaching a
closure cap onto an unused sampler.
For each sampler (one at a time), separate the primary (top) and
secondary (bottom) sections of the sampler using the edge of a coin as
a pry.
Securely snap a cup onto the bottom of the primary section.
Snap a closure cap onto the secondary section of the sampler and
assure that the attached port plugs are placed firmly into the port
holes.
Return the sampler sections with closure caps and cup in place to
the metal can which contains the PTFE tubes (which will be used by the
laboratory). Close the can with the plastic lid, and seal it with an
OSHA 21 form.
Verify that the sampling times are properly recorded on OSHA 91A
form for each sample. Also, identify blank samples on this form.
Record the room temperature and atmospheric pressure of the
sampling site on OSHA 91A form.
List any compounds that could be considered potential
interferences, especially solvents, that are being used in the
sampling area.
Submit the samples to the laboratory for analysis as soon as
possible after sampling. If delay is unavoidable, store the samples in
a refrigerator. Ship any bulk samples separate from the air
samples.
2.4 Sampler capacity (Section 4.7)
2.4.1 Charcoal tubes
The sampling capacity of the front section of a charcoal tube was
tested by sampling a dynamically generated test atmosphere of benzene
(73.4 mg/m3 or 23 ppm) at an absolute humidity of 15.7
milligrams of water per liter of air (about 80% relative humidity at
22.2°C). The samples were collected at 50 mL/min. No breakthrough was
observed, even after sampling for 600 min.
2.4.2 SKC 575-002 Passive Samplers and 3M 3520 OVMs
The sampling rate and capacity of the SKC 575-002 Passive Sampler
and the 3M 3520 OVM were determined by sampling a dynamically
generated test atmosphere of benzene (7.34 mg/m3 or 2.3 ppm
and 73.4 mg/m3 or 23 ppm) at an absolute humidity of 15.7
milligrams of water per liter of air (about 80% relative humidity at
22.2°C) for increasing time intervals. A sampling rate of 17.1 mL/min
for SKC 575-002 Passive Samplers and 34.3 mL/min for 3M 3520 OVM was
determined. The recommended sampling times for this method are 10
minutes for Peak, 15 minutes for STEL, and 240 minutes for TWA
sampling. The tests showed a significant difference in the sampling
rate for a 5 minute sample versus the determined sampling rate,
therefore, these samplers cannot be used to sample for less than 10
minutes.
2.5 Extraction efficiency (Section 4.8)
It is the responsibility of each analytical laboratory to determine
the extraction efficiency because the adsorbent material, internal
standard, reagents and laboratory techniques may be different than the
those listed in this evaluation and influence the results.
2.5.1 Charcoal tubes
The mean extraction efficiency for benzene from dry charcoal tubes
over the range of RQL to 2 times the target concentration (0.13 to
79.1 micrograms per sample) was 97.0%. The extraction efficiency was
not affected by the presence of water (average recovery of 96.7%).
Extracted samples remain stable for at least 24 h.
2.5.2 SKC 575-002 Passive Samplers
The mean extraction efficiency for benzene from dry SKC 575-002
Passive Samplers over the range of RQL to 2 times the target
concentration (0.13 to 24.6 micrograms per sample) was 93.6%. The
extraction efficiency was not affected by the presence of water
(average recovery of 93.9%).
Extracted samples remain stable for at least 24 h.
2.5.3 3M 3520 OVMs
The mean extraction efficiency for benzene from dry 3M 3520 OVMs
over the range of RQL to 2 times the target concentration (0.13 to
54.5 micrograms per sample) was 97.9%. The extraction efficiency was
not affected by the presence of water (average recovery of 98.1%).
Extracted samples remain stable for at least 24 h.
2.6 Recommended sampling time and sampling rate
2.6.1 Charcoal tubes
Sample with charcoal tubes for up to 240 min at 50 mL/min (12 L) to
collect TWA (long-term) samples, for 10 min at 50 mL/min (0.5 L) to
collect Peak (short-term) samples, and for 15 min at 50 mL/min (0.75
L) collect STEL (short-term) samples.
When short-term samples are collected, the air concentration
equivalent to the reliable quantitation limit becomes larger. For
example, the reliable quantitation limit for charcoal tubes is 0.05
ppm (0.17 mg/m3) for benzene when 0.75 L are collected.
2.6.2 SKC 575-002 Passive Samplers
Sample with SKC 575-002 Passive Samplers for up to 240 min to
collect TWA (long-term) samples, for 10 min to collect Peak
(short-term) samples, and for 15 min to collect STEL (short-term)
samples. The sampling rate is 17.1 mL/min.
When short-term samples are collected, the air concentration
equivalent to the reliable quantitation limit becomes larger. For
example, the reliable quantitation limit for SKC 575-002 Passive
Samplers is 0.20 ppm (0.64 mg/m3) for benzene when 0.26 L
(15 min) are collected.
2.6.3 3M 3520 OVMs
Sample with 3M 3520 OVMs for up to 240 min to collect TWA
(long-term) samples, for 10 min to collect Peak (short-term) samples,
and for 15 min to collect STEL (short-term) samples. The sampling rate
is 34.3 mL/min.
When short-term samples are collected, the air concentration
equivalent to the reliable quantitation limit becomes larger. For
example, the reliable quantitation limit for 3M 3520 OVMs is 0.08 ppm
(0.26 mg/m3) for benzene when 0.51 L (15 min) are
collected.
2.7 Interferences, sampling (Section 4.9)
2.7.1 Charcoal tubes
Retention
The mean retention efficiency for all samples was 100.4%, when
charcoal tubes containing 880 g of benzene were allowed to sample 9 L
of contaminant-free air having an absolute humidity of 15.7 milligrams
of water per liter of air (about 80% relative humidity at 22.2°C).
Low humidity
The ability of a charcoal tube to collect benzene from a relatively
dry atmosphere was determined by sampling an atmosphere of two times
the target concentration of benzene and having an absolute humidity of
1.9 milligrams of water per liter of air (about 10% relative humidity
at 22.2°C). The samples collected above 99.7% of theoretical.
Low concentration
The ability of a charcoal tube to collect benzene at low
concentrations was determined by sampling a test atmosphere containing
0.1 times the target concentration of benzene and having an absolute
humidity of 15.7 milligrams of water per liter of air (about 80%
relative humidity at 22.2°C). The samples collected above 98.5% of
theoretical.
Interference
The ability of charcoal tubes to collect benzene in the presence of
an interference was determined from a test atmosphere containing one
times the target concentration of benzene, 890 mg/m3 of
gasoline, and having an absolute humidity of 15.7 milligrams of water
per liter of air (about 80% relative humidity at 22.2°C). The benzene
concentration on the samples remained above 98.9% of theoretical.
2.7.2 SKC 575-002 Passive Samplers
Reverse diffusion
Reverse diffusion is the measure of the ability of the sorbent
within a diffusive sampler to retain the analyte collected. Reverse
diffusion is measured by first exposing two sets of samplers to humid
air containing the analyte, and then additionally exposing one of the
sets to clean humid air of an absolute humidity of 15.7 milligrams of
water per liter of air (about 80% relative humidity at 22.2°C).
Comparison of the two sets of SKC 575-002 Passive Samplers showed that
an average of 99.1% of the benzene was retained, indicating a loss of
0.9% to reverse diffusion. The loading of benzene on the samplers was
75.3 µg. (Section 4.9.2)
Low humidity
The ability of a SKC 575-002 Passive Sampler to collect benzene
from a relatively dry atmosphere was determined by sampling an
atmosphere of two times the target concentration of benzene and having
an absolute humidity of 1.9 milligrams of water per liter of air
(about 10% relative humidity at 22.2°C). The samples collected above
98.9% of theoretical.
Low concentration
The recovery for all samples was above 98.7% of theoretical, when
SKC 575-002 Passive Samplers were used to sample a test atmosphere
containing 0.1 times the target concentration of benzene and having an
absolute humidity of 15.7 milligrams of water per liter of air (about
80% relative humidity at 22.2°C).
Interference
The ability of SKC 575-002 Passive Samplers to collect benzene in
the presence of an interference was determined from a test atmosphere
containing one times the target concentration of benzene, 890
mg/m3 of gasoline, and having an absolute humidity of 15.7
milligrams of water per liter of air (about 80% relative humidity at
22.2°C). The benzene concentration on the samples remained above 99.3%
of theoretical.
2.7.3 3M 3520 OVMs
Reverse diffusion
Reverse diffusion is the measure of the ability of the sorbent
within a diffusive sampler to retain the analyte collected. Reverse
diffusion is measured by first exposing two sets of samplers to humid
air containing the analyte, and then additionally exposing one of the
sets to clean humid air of an absolute humidity of 15.7 milligrams of
water per liter of air (about 80% relative humidity at 22.2°C).
Comparison of the two sets of 3M 3520 OVMs showed that an average of
99.2% of the benzene was retained, indicating a loss of 0.8% to
reverse diffusion. The loading of benzene on the samplers was 75.3 µg.
(Section 4.9.3)
Low humidity
The ability of 3M 3520 OVMs to collect benzene from a relatively
dry atmosphere was determined by sampling an atmosphere of two times
the target concentration of benzene and having an absolute humidity of
1.9 milligrams of water per liter of air (about 10% relative humidity
at 22.2°C). The samples collected above 98.7% of theoretical.
Low concentration
The ability of 3M 3520 OVMs to collect benzene at low
concentrations was determined by sampling a test atmosphere containing
0.1 times the target concentration of benzene and having an absolute
humidity of 15.7 milligrams of water per liter of air (about 80%
relative humidity at 22.2°C). The samples collected above 99.5% of
theoretical.
Interference
The ability of 3M 3520 OVMs to collect benzene in the presence of
an interference was determined from a test atmosphere containing one
times the target concentration of benzene, 890 mg/m3 of
gasoline, and having an absolute humidity of 15.7 milligrams of water
per liter of air (about 80% relative humidity at 22.2°C). The benzene
concentration on the samples remained above 99.1% of
theoretical.
- Analytical Procedure
Adhere to the rules set down in your Chemical Hygiene
Plan16. Avoid skin contact and inhalation of all chemicals
and review all MSDSs before beginning this analytical
procedure.
3.1 Apparatus
3.1.1 Gas chromatograph equipped with an FID. A Hewlett-Packard
Model 5890 Series II GC equipped with an integrator, an automatic
sample injector, and an FID was used in this evaluation.
3.1.2 A GC column capable of separating benzene from the extracting
solvent, internal standard, and the components of gasoline. A J&W
60-m × 0.32-mm i.d. DB-1 (5-µm df) capillary column was used in this
evaluation.
3.1.3 An electronic integrator or other suitable means of measuring
GC detector response. A Waters Millenium32 Data System was
used in this evaluation, along with a Hewlett Packard 3396 Series II
integrator.
3.1.4 Glass vials with PTFE-lined caps. For this evaluation 2 and
4-mL vials were used.
3.1.5 A dispenser capable of delivering 1.0 or 2.0 mL of extracting
solvent to prepare standards and samples. If a dispenser is not
available, 1.0- and 2.0-mL volumetric pipets may be used.
3.1.7 Volumetric flasks - 10-mL and other convenient sizes for
preparing standards.
3.1.8 Calibrated 10-L syringe for preparing standards.
3.1.9 An SKC Desorption shaker with rack (226D-03K) was used to
extract SKC 575-002 Passive Samplers in this evaluation.
3.1.10 A mechanical shaker. An Eberbach mechanical shaker was used
to extract the charcoal tubes in this evaluation.
3.2 Reagents
3.2.1 Benzene, [CAS no. 71-43-2], reagent grade or better. The
benzene used in this evaluation was A.C.S. reagent grade (lot no. CU
03251PS) purchased from Aldrich (Milwaukee, WI).
3.2.2 Carbon disulfide (CS2), [CAS no. 75-15-0], reagent
grade or better. The carbon disulfide used in this evaluation was
99.9+% low benzene content grade (lot no. TI 01762PI) purchased from
Aldrich (Milwaukee, WI).
3.2.3 1-Phenylhexane (n-hexylbenzene) [CAS no. 1077-16-3], reagent
grade or better. The 1-phenylhexane used in this evaluation was 97%
reagent grade (lot no. 03006PZ) purchased from Aldrich (Milwaukee,
WI).
3.2.4 The extraction solvent used for this evaluation consisted of
0.25 L/mL n-hexylbenzene (1-phenylhexane) in the CS2. The
n-hexylbenzene was added to the CS2 as an internal
standard. Other internal standards can be used provided they are fully
tested.
3.3 Standard preparation
3.3.1 Prepare concentrated stock standards of benzene in the
extracting solvent. At least two separate stock standard should be
prepared. Prepare working analytical standards by diluting these stock
standards with the extracting solution delivered from the same
dispenser used to extract the samples. For example, to prepare a
target standard (1 ppm), inject 4.5 L of benzene in a 10-mL volumetric
flask containing the extracting solvent and then make a 1/10 dilution
with the extracting solvent to obtain the working standard at the
target level. A second set of standards from a different primary
standard should be prepared to check the quality of the first set of
standards.
3.3.2 Bracket sample concentrations with standard concentrations.
If upon analysis, sample concentrations fall outside the range of
prepared standards, prepare and analyze additional standards to
confirm instrument response, or dilute high samples with extraction
solvent and reanalyze the diluted samples.
3.4 Sample preparation
3.4.1 Charcoal tubes
Remove the plastic end caps from the sample tube and carefully
transfer each section of the adsorbent to separate 2-mL vials. Discard
the glass tube and glass wool and polyurethane plugs.
Add 1.0 mL of extracting solution to each vial and immediately seal
the vials with PTFE-lined caps.
Shake the vials on a shaker for 30 min (Shaking is necessary to
obtain the extraction efficiency found in this method; without shaking
the extraction efficiencies will be lower.)
3.4.2 SKC 575-002 Passive Samplers (In general, follow the
manufacturer's instructions.)
Cut off the ends of the two protruding tubes of each sampler with a
razor blade or sharp knife.
Slowly add 2.0 mL of extraction solvent through one of the
protruding tubes (ports), stopping at least once to allow the bubbling
to subside before adding the rest of the extraction solvent.
Immediately insert plugs into the ports.
Mount the samplers in the sampler rack (SKC Cat. No. 226-04-5) of a
specialized shaker (SKC Cat. No. 226D-03-1) and shake the samplers for
1 hour.
Do not leave the extracted sample in the sampler. Transfer each
extracted sample by removing the plugs from the sampler ports, firmly
inserting the tapered end of a supplied PTFE tube into the outer port
and carefully pouring the solution through the PTFE tube into a
labeled autosampler vial. Immediately cap each vial.
3.4.3 3M 3520 OVMs (In general, follow the manufacturer's
instructions.)
Remove both sampler sections from the metal cans, along with the
sections of PTFE tubing. Assure that the closure caps are firmly
snapped to the primary and secondary sections of all the samplers.
Also assure that all cap plugs are firmly seated in the cap ports. Any
deviations must be noted. Make sure each section of the sampler is
labeled properly for future reference.
Prepare one section of sampler at time by temporarily removing the
cap plugs from the ports and adding 2.0 mL of extraction solvent
through the center port. Immediately replace the plugs in the ports.
Repeat the process for the second section.
Allow the sampler sections to extract for 30 min. Periodically
apply gentle agitation to the sampler sections during the extraction
period.
Do not leave the extracted sample in the sampler. Transfer the
solution from each sampler section by removing both plugs from the
ports, inserting a decanting spout (a small section of PTFE tubing)
into the rim port and pouring the liquid through the spout into a
labeled autosampler vial. Immediately cap each vial.
3.5 Analysis
3.5.1 Analytical conditions
| GC conditions |
|
|
column temperature:
zone
temperatures:
run time: column gas flow: septum
purge: injection size: column:
retention
times:
FID conditions hydrogen
flow: air flow: nitrogen makeup flow: |
initial 60°C, hold 5 min, program
at 10°/min to 220°C, hold 14 min
220°C
(injector) 240°C (detector) 35 min 2.4 mL/min
(hydrogen) 3.5 mL/min (hydrogen) 1.0 µL (19:1
split) 60-m × 0.32-mm i.d. capillary DB-1 (df = 5.0
µm) 9.26 min (carbon disulfide) 13.66 min
(benzene) 32.491 min (n-hexyl benzene)
35
mL/min 450 mL/min
35 mL/min |
Figure
3.5.1. Chromatogram obtained at the target concentration
with the recommended conditions. (1 = carbon disulfide; 2
= benzene; and 3 = n-hexylbenzene)
|
3.5.2 An internal standard (ISTD) calibration method is used. The
calibration curve was constructed by plotting the ISTD-corrected
response of standard injections versus micrograms of analyte per
sample. Bracket the samples with freshly prepared analytical standards
over a range of concentrations.
 |
| Figure 3.5.2. A calibration curve of
benzene. (Y = 1356X +
199) |
3.6 Interferences (analytical)
3.6.1 Any compound that produces an FID response and has a similar
retention time as the analyte or internal standard is a potential
interference. If any potential interferences were reported, they
should be considered before samples are extracted. Generally,
chromatographic conditions can be altered to separate an interference
from the analyte.
3.6.2 When necessary, the identity of an analyte peak may be
confirmed with additional analytical data (Section
4.10).
3.7 Calculations
3.7.1 Charcoal tubes
The amount of benzene per sampler is obtained from the appropriate
calibration curve in terms of micrograms per sample, uncorrected for
extraction efficiency. The back section is analyzed primarily to
determine the extent of sampler saturation. (The charcoal tube is
considered saturated when 25% of the amount found on the front section
is found on the back section of the tube, and therefore, some of the
sample may have been lost.) If any analyte is found on the back
section, it is added to the amount on the front section. This total
amount is then corrected by subtracting the total amount (if any)
found on the blank. The air concentration is calculated using the
following formulas.
 |
where |
CM is
concentration by weight (mg/m3) |
|
M is micrograms per
sample |
|
V is liters of air
sampled |
|
EE is extraction
efficiency, in decimal form |
 |
where |
CV is concentration by
volume (ppm) |
|
VM = 24.46 at NTP |
|
CM is concentration by
weight |
|
Mr is molecular weight of
78.11 |
3.7.2 3M 3520 OVMs and SKC 575-002 Passive Samplers
The amount of benzene for the samples is obtained from the
appropriate calibration curve in terms of micrograms per sample,
uncorrected for extraction efficiency. (In the case of the 3M 3520
OVMs, the back section is analyzed primarily to determine the extent
of sampler saturation. If any analyte is found on the back section,
the amount is multiplied by 2.2 (as per manufacturer's instructions)
and then added to the amount on the front section. The sampler is
saturated, affecting its ability to collect, when the corrected amount
found on the back section is 50% of the amount found on the front
section.) This total amount is then corrected by subtracting the total
amount (if any) found on the blank. The air concentration is
calculated using the following formulas.
 |
where |
RSS is the sampling rate at
sampling site |
|
RNTP is the
sampling rate at NTP conditions (SKC 575-002 = 17.1 mL/min, 3M
OVM = 34.3 mL/min) |
|
TSS is the
sampling site temperature in K |
|
TNTP is 298.2
K |
|
PSS is the
sampling site pressure in mmHg |
|
PNTP is 760
mmHg |
 |
where |
CM is concentration by
weight (mg/m3) |
|
M is micrograms per sample |
|
RSS is the sampling rate at
the sampling site |
|
t is the sampling time |
|
EE is extraction efficiency,
in decimal form |
 |
where |
CV is concentration by
volume (ppm) |
|
VM = 24.46 at NTP |
|
CM is concentration by
weight |
|
Mr is molecular weight of
78.11 |
If the sampling site temperature is not provided, assume that it is
22.2°C. If the sampling site atmospheric pressure is not given,
calculate an approximate value based on the sampling site elevation
from the following equation.
 |
where |
PSS is the approximate
atmospheric pressure |
|
E is the sampling site elevation,
ft |
|
A is 3.887×10-7
mmHg/ft2 |
|
B is 0.02748
mmHg/ft |
- Backup Data
General background information about the determination of detection
limits and precision of the overall procedure is found in the "Evaluation
Guidelines for Air Sampling Methods Utilizing Chromatography
Analysis"17. The Guidelines define analytical parameters,
specify required laboratory tests, statistical calculations and acceptance
criteria.
4.1 Detection limit of the analytical procedure (DLAP)
The DLAP is measured as the mass of analyte introduced onto the
chromatographic column. Ten analytical standards were prepared with
equal increments with the highest standard containing 1050 ng/mL. This
is the concentration that would produce a peak approximately 10 times
the response of a reagent blank near the elution time of the analyte.
These standards, and the reagent blank were analyzed with the
recommended analytical parameters (1-µL injection with a 19:1 split),
and the data obtained were used to determine the required parameters
(standard error of estimate and slope) for the calculation of the DLAP.
Values of 18.0 and 18.74 were obtained for the slope and standard error
of estimate respectively. DLAP was calculated to be 3.12 pg.
|
Table 4.1 Detection Limit of the Analytical
Procedure |
|
|
Figure 4.1. Plot of data to determine the
DLAP. (Y = 18.0X + 22.8) |
| concentration (ng/mL) |
mass on column (pg) |
area counts (µV-s) |
|
|
0 105 211 316 422 527 633 738 844 949 1050 |
0 5.5 11.1 16.6 22.2 27.7 33.3 38.8 44.4 49.9 55.3 |
0 111 249 341 402 545 618 722 835 924 1001 |
|
4.2 Detection limit of the overall procedure (DLOP) and reliable
quantitation limit (RQL)
DLOP is measured as mass per sample and expressed as equivalent air
concentrations, based on the recommended sampling parameters. Ten
samplers were spiked with equal descending increments of analyte, such
that the highest sampler loading was 1050 ng/sample. This is the amount
spiked on a sampler that would produce a peak approximately 10 times the
response of a sample blank. These spiked samplers, and the sample blank
were analyzed with the recommended analytical parameters, and the data
obtained used to calculate the required parameters (standard error of
estimate and the slope) for the calculation of the DLOP. Values of 0.940
and 11.90 were obtained for the slope and standard error of estimate for
charcoal tubes, respectively. The DLOP for charcoal tubes was calculated
to be 38 ng/sample (1 ppb). Values of 0.958 and 15.58 were obtained for
the slope and standard error of estimate for SKC 575-002 Passive
Samplers, repsectively. The DLOP for SKC 575-002 Passive Samplers was
calculated to be 49 ng/sample (4 ppb). Values of 0.918 and 12.45 were
obtained for the slope and standard error of estimate for 3M 3520 OVMs,
respectively. The DLOP for 3M 3520 OVMs was calculated to be 41
ng/sample (1 ppb).
|
Table 4.2.1 Detection Limit of the Overall
Procedure on Charcoal Tubes |
|
|
Figure 4.2.1. Plot of data to determine
the DLOP/RQL on charcoal tubes (lot 2000). (Y = 0.940X -
8.80) |
| mass per sample (ng) |
area counts (µV-s) |
|
0 105 211 316 422 527 633 738 844 949 1050 |
0 103 175 293 383 488 563 683 784 889 992 |
|
|
Table 4.2.2 Detection Limit of the Overall
Procedure on SKC 575-002 Passive Samplers |
|
|

Figure 4.2.2. Plot of data to determine
the DLOP/RQL on SKC 575-002 Passive Samplers. (Y = 0.958X
+ 30.0) |
| mass per sample (ng) |
area counts (µV·s) |
|
0 105 211 316 422 527 633 738 844 949 1050 |
0 114 255 346 409 551 623 728 841 944 1040 |
|
|
Table 4.2.3 Detection Limit of the Overall
Procedure on 3M 3520 OVMs |
|
|

Figure 4.2.3. Plot of data to determine
the DLOP/RQL on 3M 3520 OVMs. (Y = 0.918X + 11.5)
|
| mass per sample (ng) |
area counts (µV·s) |
|
0 105 211 316 422 527 633 738 844 949 1050 |
0 111 211 296 391 517 606 687 786 862 982 |
|
The RQL is considered the lower limit for precise quantitative
measurements. It is determined from the regression line parameters
obtained for the calculation of the DLOP, providing 75% to 125% of the
analyte is recovered. The RQLs for the various media are listed in
Table 4.2.4.
Table 4.2.4 Reliable Quantitation Limits
|
| sampler |
ng |
ppb |
µg/m3 |
EE |
|
charcoal tube SKC
575-002 3M 3520 |
127 163 136 |
3.32 4.26 3.55 |
10.6 13.6 11.3 |
96.6 93.1 97.9 |
|
 |
| Figure 4.2.4. Chromatogram of the
RQL. (1 = benzene) |
4.3 Instrument calibration
The standard error of estimate was determined from the linear
regression of data points from standards over a range that covers 0.25
to 2 times the TWA target concentration. A calibration curve was
constructed and shown in Section 3.5.2 from the six injections of five
standards. The standard error of estimate is 0.34 µg.
Table 4.3 Instrument Calibration
|
| standard concn µg/mL |
area
counts (µV·s) |
|
9.9 19.8 39.5 59.3 79.1 |
14230 26684 53374 80390 107770 |
14187 26783 53487 80450 108450 |
14329 26645 53380 80334 107689 |
14284 26743 53409 80299 107651 |
14229 26523 53312 80387 107743 |
14219 26589 53287 80401 107812 |
|
4.4 Precision (overall procedure)
4.4.1 Charcoal tubes
The precision at the 95% confidence level is obtained by
multiplying the standard error of estimate by 1.96 (the z-statistic
from the standard normal distribution at the 95% confidence level). In
Section 4.5, 95% confidence intervals are drawn about their respective
regression lines in the storage graph figures. The precision of the
overall procedure of ±9.94 % was obtained from the standard error of
estimate of 5.07% in Figure 4.5.1.1. The precision includes an
additional 5% for sampling error.
4.4.2 SKC 575-002 Passive Samplers
The precisions of the overall procedure at the 95% confidence level
for the ambient temperature 19-day storage test (at the target
concentration) from SKC 575-002 Passive Samplers are given in Table
4.4.2. They each include an additional 8.7% for sampling rate
variability. There are different values given, depending on whether
both, either, or neither temperature (T) or atmospheric
pressure (P) are known at the sampling site. If the sampling
site temperature is unknown, it is assumed to be 22.2 ± 15°C (72 ±
27°F) and a variability of ±7.7% is included. If the atmospheric
pressure is not known, it is estimated from the sampling site
elevation and a variability of ±3% is included.
Table 4.4.2 Standard Error of Estimate and Precision of
the Overall Procedure for SKC 575-002 Passive
Samplers
|
| know condition |
error (%) |
precision (±%) |
|
both T &
P only T only P neither T
nor P |
8.75 9.25 11.65 12.03 |
17.2 18.1 22.8 23.6 |
|
4.4.3 3M 3520 OVMs
The precisions of the overall procedure at the 95% confidence
level for the ambient temperature 19-day storage test (at the target
concentration) from 3M 3520 OVMs are given in Table 4.4.3. They each
include an additional 6.4% for sampling rate variability. There are
different values given, depending on whether both, either, or neither
temperature (T) or atmospheric pressure (P) are
known at the sampling site. If the sampling site temperature is
unknown, it is assumed to be 22.2 ± 15°C (72 ± 27°F) and a variability
of ±7.7% is included. If the atmospheric pressure is not known, it is
estimated from the sampling site elevation and a variability of ±3% is
included.
Table 4.4.3 Standard Error of Estimate and
Precision of the Overall Procedure for 3M 3520
OVMs
|
| know condition |
error (%) |
precision (±%) |
|
both T &
P only T only P neither T
nor P |
6.45 7.11 10.0 10.5 |
12.6 13.9 19.6 20.6 |
|
4.5 Storage test
4.5.1 Charcoal tubes
Storage samples for benzene were prepared by collecting samples
from a controlled test atmosphere using the recommended sampling
conditions. The concentration of benzene was at the target
concentration and the absolute humidity was 15.7 milligrams of water
per liter of air (about 80% at 22.2°C). Thirty-three storage samples
were prepared. Three samples were analyzed on the day of generation.
Fifteen of the tubes were stored at reduced temperature (4°C) and the
other fifteen were stored in a closed drawer at ambient temperature
(about 22°C). At 2-5 day intervals, three samples were selected from
each of the two storage sets and analyzed. Sample results are not
corrected for extraction efficiency.
Table 4.5.1 Storage Test for Benzene on Charcoal
Tubes
|
| time (days) |
ambient storage
recovery (%) |
refrigerated
storage recovery (%) |
|
0 5 8 12 15 19 |
98.9 99.4 97.9 98.1 98.5 96.8 |
99.9 98.9 99.4 99.9 98.9 99.1 |
100.4 97.9 99.1 97.9 96.9 98.4 |
98.9 98.8 99.5 100.1 97.7 97.9 |
99.9 100.3 100.4 99.9 99.1 99.8 |
100.4 99.7 98.8 97.9 98.5 98.4 |
|
 |
| Figure 4.5.1.1. Ambient storage for benzene
collected on charcoal tubes. |
 |
| Figure 4.5.1.2. Refrigerated storage test for benzene
collected on charcoal tubes. |
4.5.2 SKC 575-002 Passive Samplers
Storage samples for benzene were prepared by collecting samples
from a controlled test atmosphere using the recommended sampling
conditions. The concentration of benzene was at the target
concentration and the absolute humidity was 15.7 milligrams of water
per liter of air (about 80% at 22.2°C). Thirty-three storage samples
were prepared. Three samples were analyzed on the day of generation.
Fifteen of the samplers were stored at reduced temperature (4°C) and
the other fifteen were stored in a closed drawer at ambient
temperature (about 22°C). At 2-5 day intervals, three samples were
selected from each of the two storage sets and analyzed. Sample
results are not corrected for extraction efficiency.
Table 4.5.2 Storage Test for Benzene on SKC 575-002
Passive Samplers
|
| time (days) |
ambient
storage recovery (%) |
refrigerated
storage recovery (%) |
|
0 5 8 12 15 19 |
100.5 99.6 99.4 97.9 96.6 97.5 |
100.6 100.1 98.8 98.8 98.6 96.5 |
98.9 98.8 96.8 98.2 96.1 95.8 |
100.5 99.3 99.9 100.1 98.9 99.3 |
100.6 99.4 99.1 98.6 99.6 97.5 |
98.9 100.3 98.8 99.5 97.5 99.1 |
|
 |
| Figure 4.5.2.1. Ambient storage for
benzene collected on SKC 575-002 Passive
Samplers. |
 |
| Figure 4.5.2.2. Refrigerated storage test for
benzene collected on SKC 575-002 Passive
Samplers. |
4.5.3 3M 3520 OVMs
Storage samples for benzene were prepared by collecting samples
from a controlled test atmosphere using the recommended sampling
conditions. The concentration of benzene was at the target
concentration and the absolute humidity was 15.7 milligrams of water
per liter of air (about 80% at 22.2°C). Thirty-three storage samples
were prepared. Three samples were analyzed on the day of generation.
Fifteen of the samplers were stored at reduced temperature (4°C) and
the other fifteen were stored in a closed drawer at ambient
temperature (about 22°C). At 2-5 day intervals, three samples were
selected from each of the two storage sets and analyzed. Sample
results are not corrected for extraction efficiency.
Table 4.5.3 Storage Test for Benzene on 3M 3520
OVMs
|
| time (days) |
ambient
storage recovery (%) |
refrigerated storage
recovery (%) |
|
0 5 8 12 15 19 |
100.1 98.6 99.1 96.2 97.2 96.6 |
99.2 99.1 97.8 98.9 95.9 95.5 |
98.2 97.8 98.3 97.1 97.2 95.9 |
100.1 99.9 98.6 99.1 98.5 99.3 |
99.2 98.2 98.9 97.6 99.2 98.5 |
98.2 97.9 96.8 97.8 97.2 97.3 |
|
 |
| Figure 4.5.3.1. Ambient storage for
benzene collected on 3M 3520
OVMs. |
 |
| Figure 4.5.3.2. Refrigerated storage test for
benzene collected on 3M 3520
OVMs. |
4.6 Reproducibility
Six samples were prepared for the three types of samplers by
collecting them from a controlled test atmosphere similar to that which
was used in the collection of the storage samples. The samples were
submitted to the OSHA Salt Lake Technical Center for analysis, along
with a draft copy of this method. The samples were analyzed after being
stored for 7 days at 4°C. Sample results were corrected for extraction
efficiency. No sample result for benzene had a deviation greater than
the precision of the overall procedure determined in Section 4.4.
Table 4.6.1 Reproducibility Data for Benzene on Charcoal
Tubes
|
| theoretical (µg/sample) |
recovered (µg/sample) |
recovery (%) |
deviation (%) |
|
76.91 83.34 77.84 88.56 |
73.61 81.58 75.17 85.62 |
95.7 97.9 96.6 96.7 |
-4.3 -2.1 -3.4 -3.3 |
|
Table 4.6.2 Reproducibility Data for Benzene on
SKC 575-002 Passive Samplers
|
| theoretical (µg/sample) |
recovered (µg/sample) |
recovery (%) |
deviation (%) |
|
26.38 26.38 26.38 26.38 26.38 |
25.62 25.03 25.43 25.21 26.01 |
97.1 94.9 96.4 95.6 98.5 |
-2.9 -5.1 -3.6 -4.4 -1.5 |
|
Table 4.6.3 Reproducibility Data for Benzene
on 3M 3520 OVMs
|
| theoretical (µg/sample) |
recovered (µg/sample) |
recovery (%) |
deviation (%) |
|
58.52 58.52 58.52 58.52 58.52 58.52 |
56.53 56.18 55.95 57.06 55.59 57.01 |
96.6 96.0 95.6 97.5 95.0 97.4 |
-3.4 -4.0 -4.4 -2.5 -5.0 -2.6 |
|
4.7 Sampler capacity
4.7.1 Charcoal tubes
The sampling capacity of the front section of a charcoal tube was
tested by sampling from a dynamically generated test atmosphere of
benzene (72.8 mg/m3 or 23 ppm) with an absolute humidity of
15.7 milligrams of water per liter of air (about 80% relative humidity
at 22.2°C). This air concentration was twice the highest PEL. The
samples were collected along with diffusive samplers. Three charcoal
tubes and three of each kind of diffusive samplers were collected at
5, 10, 15, and 30 min, and 1, 2, 3, 4, 6, 8, and 10 hours. The
charcoal tube samples were collected at 50 mL/min. There was no
breakthrough observed in any of the charcoal tubes, therefore the
sampler capacity was never exceeded. The interference study of
gasoline showed that for samples taken up to 8 hours, with a benzene
concentration of 32 mg/m3 or 10 ppm, and a gasoline
concentration of 890 mg/m3 had no breakthrough for the
benzene, though the gasoline did breakthrough after 4 hours (Section
4.9).
4.7.2 SKC 575-002 Passive Samplers
The sampling rate and sampler capacity are determined with samples
collected for increasing time intervals from a controlled test
atmosphere. Sampler capacity is exceeded when the sampling rate
decreases (greater than 10 hours for SKC 575-002). The concentration
of the test atmosphere was two times the target concentration with an
absolute humidity of 15.7 milligrams of water per liter of air (about
80% at 22.2°C). The preliminary sampling rate was determined by
averaging the nine values for the 0.5, 1 and 2 h samples. Horizontal
lines were placed 10% above and below the preliminary sampling rate.
For an atmosphere of 2.3 ppm the sampling rate is 17.2 mL/min at 760
mmHg and 25°C and represents the average of all values between the
lines. The standard deviation and RSD are 0.453 mL/min and 2.63%,
respectively. For an atmosphere of 23 ppm the sampling rate is 16.9
mL/min at 760 mmHg and 25 °C and represents the average of all values
between the lines. The standard deviation and RSD are 0.411 mL/min and
2.43%, respectively. The average sampling rate from both
determinations was 17.1 mL/min. The data obtained are shown in Table
4.7.2.1 and Figure 4.7.2.1. Mass collected is corrected for extraction
efficiency. The sampling rate for a 5-minute sample is significantly
different from the calculated sampling rate indicating that samples
should not be taken for that short of a time. The recommended sampling
time is 4 h for TWA samples, 10 min for Peak samples, and 15 min for
STEL samples.
Table 4.7.2.1 Determination of Sampling Rate and
Time for SKC 575-002 Passive Samplers from a 2.3-ppm
Atmosphere
|
|
sampling rate
(mL/min) |
| time (h) |
first |
second |
third |
|
0.083 0.167 0.25 0.5 1 2 3 4 6 8 10 |
19.2 18.3 18.0 17.8 17.2 16.9 16.2 16.3 16.4 16.2 16.0 |
19.0 18.6 17.9 17.5 17.3 16.8 16.8 16.1 16.2 16.5 16.1 |
18.8 18.4 18.1 17.7 17.4 16.4 16.3 16.0 16.3 16.1 15.9 |
|
 |
| Figure 4.7.2.1. The SKC 575-002 Passive
Sampler data plotted to determine the recommended sampling time
and sampling rate from a 2.3-ppm
atmosphere. |
Table 4.7.2.2 Determination of Sampling Rate and Time for
SKC 575-002 Passive Samplers from a 23-ppm Atmosphere
|
|
sampling rate
(mL/min) |
| time (h) |
first |
second |
third |
|
0.083 0.167 0.25 0.5 1 2 3 4 6 8 10 |
18.4 18.0 17.4 17.2 16.7 16.5 16.1 16.4 16.4 16.4 16.3 |
18.5 17.7 17.7 17.4 16.5 16.3 16.3 16.1 16.3 16.1 16.0 |
18.2 17.9 17.6 16.9 16.9 16.9 16.5 16.6 16.2 16.0 16.1 |
|
 |
| Figure 4.7.2.2. The SKC 575-002 Passive
Sampler data plotted to determine the recommended sampling time
and sampling rate from a 23-ppm
atmosphere. |
4.7.3 3M 3520 OVMs
The sampling rate and sampler capacity are determined with samples
collected for increasing time intervals from a controlled test
atmosphere. Sampler capacity is exceeded when the sampling rate
decreases(greater than 10 hours for 3M 3520 OVMs). The concentration
of the test atmosphere was two times the target concentration with an
absolute humidity of 15.7 milligrams of water per liter of air (about
80% at 22.2°C). The preliminary sampling rate was determined by
averaging the nine values for the 0.5, 1 and 2 h samples. Horizontal
lines were placed 10% above and below the preliminary sampling rate.
For an atmosphere of 2.3 ppm the sampling rate is 34.3 mL/min at 760
mmHg and 25°C and represents the average of all values between the
lines. The standard deviation and RSD are 0.771 mL/min and 2.25%,
respectively. The data obtained are shown in Table 4.7.3.1 and Figure
4.7.3.1. For an atmosphere of 23 ppm the sampling rate is 34.3 mL/min
at 760 mmHg and 25 °C and represents the average of all values between
the lines. The standard deviation and RSD are 0.466 mL/min and 1.36%,
respectively. The data obtained are shown in Table 4.7.3.2 and Figure
4.7.3.2. Mass collected is corrected for extraction efficiency. The
recommended sampling time is 4 h for TWA samples, 10 min for Peak
samples, and 15 min for STEL samples.
Table 4.7.3.1 Determination of Sampling Rate and Time for
3M 3520 OVMs from a 2.3-ppm Atmosphere
|
|
sampling rate
(mL/min) |
| time (h) |
first |
second |
third |
|
0.083 0.167 0.25 0.5 1 2 3 4 6 8 10 |
36.2 35.6 35.1 34.4 34.5 33.5 34.1 34.2 32.6 34.5 34.3 |
36.3 35.9 35.8 35.9 34.1 33.6 33.5 34.4 33.9 33.2 33.8 |
36.6 35.8 35.4 34.9 33.8 33.7 33.2 34.1 33.7 33.9 33.9 |
|
|
| Figure 4.7.3.1. The 3M 3520 OVM data
plotted to determine the recommended sampling time and sampling
rate from a 2.3-ppm
atmosphere. |
Table 4.7.3.2 Determination of Sampling Rate and Time for
3M 3520 OVMs from a 23-ppm Atmosphere
|
|
sampling rate
(mL/min) |
| time (h) |
first |
second |
third |
|
0.083 0.167 0.25 0.5 1 2 3 4 6 8 10 |
36.8 35.5 35.2 34.0 34.1 33.9 33.9 33.8 34.1 34.3 34.2 |
36.0 35.0 35.6 35.2 34.4 34.5 34.2 33.9 33.7 33.8 33.6 |
36.4 35.7 35.3 34.7 33.9 34.2 33.8 34.5 34.3 34.8 33.8 |
|
 |
| Figure 4.7.3.2. The 3M 3520 OVM data
plotted to determine the recommended sampling time and sampling
rate from a 23-ppm
atmosphere. |
4.8 Extraction efficiency and stability of extracted samples
The extraction efficiency is dependent on the extraction solvent as
well as the internal standard. The extraction solvent used for this
evaluation consisted of 0.25 µL/mL n-hexylbenzene (1-phenylhexane) in
the CS2. Other extraction solvents or internal standards may
be used provided that the new extraction solution or internal standard
is tested. The new extraction solvent or internal standard should be
tested as described below.
4.8.1 Charcoal tubes
Extraction efficiency
The extraction efficiencies of benzene were determined by
liquid-spiking four charcoal tubes, at each concentration level, with
the analyte from the RQL to 2 times the target concentration. These
samples were stored overnight at ambient temperature and then
analyzed. The mean extraction efficiency over the working range of the
RQL to 2 times the target concentration is 97.0%. The extraction
efficiency for the wet samplers was not included in the overall mean
because it would bias the results. The test of wet samplers was
performed to determine if the amount of water, that would collect
under high humidity conditions at the recommended air volume, would
affect the extraction efficiency.
Table 4.8.1.1 Extration Efficiency of Benzene from
Charcoal Tubes
|
| level |
sample
number |
|
| × target concn |
µg per sample |
1 |
2 |
3 |
4 |
mean |
|
RQL 0.25 0.5 1.0 1.5 2.0
1.0
(wet) |
0.13 9.9 19.8 39.5 59.3 79.1
39.5 |
98.5 98.6 97.6 96.7 96.9 97.2
97.1 |
97.3 95.3 98.2 97.1 95.5 96.5
95.6 |
95.3 96.6 97.6 96.3 96.9 97.2
96.8 |
95.1 97.3 98.2 97.0 97.3 97.1
97.2 |
96.6 97.0 97.9 96.8 96.7 97.0
96.7 |
|
Stability of extracted samples
The stability of extracted samples was investigated by reanalyzing
the target concentration samples 24 h after initial analysis. After
the original analysis was performed two vials were recapped with new
septa while the remaining two retained their punctured septa. The
samples were reanalyzed with fresh standards. The average percent
change was 0.9% for samples that were resealed with new septa and 3.4%
for those that retained their punctured septa. Each septum was
punctured 5 times for each injection. The test was performed at room
temperature.
Table 4.8.1.2 Stability Samples for Benzene on Charcoal
Tubes
|
| punctured septa
replaced |
punctured septa
retained |
initial (%) |
after one day (%) |
difference (%) |
initial (%) |
after one day (%) |
difference (%) |
|
96.7 97.1
96.9 |
95.6 96.3 (mean) 96.0 |
-1.1 -0.8
-0.9 |
96.3 97.0
96.7 |
92.1 94.5 (mean) 93.3 |
-4.2 -2.5
-3.4 |
|
4.8.2 SKC 575-002 Passive Sampler
Extraction efficiency
The extraction efficiencies of benzene were determined by
liquid-spiking four SKC 575-002 Passive Samplers, at each
concentration level, with the analyte at the RQL to 2 times the target
concentration. These samples were stored overnight at ambient
temperature and then extracted and analyzed. The average extraction
efficiency over the working range of RQL to 2 times the target
concentration was 93.6%. The extraction efficiency for the wet
samplers was not included in the overall mean because it would bias
the results. The test of wet samplers was performed to determine if
the amount of water, that would collect under high humidity conditions
at the recommended air volume, would affect the extraction
efficiency.
Table 4.8.2.1 Extraction Efficiency of
Benzene from SKC 575-002 Passive Samplers
|
| level |
sample number |
|
| × target concn |
µg per sample |
1 |
2 |
3 |
4 |
mean |
|
RQL 0.25 0.5 1.0 1.5 2.0
1.0 (wet) |
0.13 3.08 6.15 12.3 18.5 24.6
12.3 |
93.8 95.2 93.3 94.9 92.4 94.2
93.5 |
92.4 94.3 94.3 93.6 93.8 93.4
94.5 |
91.4 93.4 93.8 94.3 93.5 94.2
93.6 |
94.6 93.1 92.6 92.2 93.3 93.1
94.0 |
93.1 94.0 93.5 93.8 93.3 93.7
93.9 |
|
Stability of extracted samples
The stability of extracted samples was investigated by reanalyzing
the target concentration samples 24 h after initial analysis. After
the original analysis was performed two vials were recapped with new
septa while the remaining two retained their punctured septa. The
samples were reanalyzed with fresh standards. The average percent
change was 1.3% for samples that were resealed with new septa and 2.6%
for those that retained their punctured septa. Each septum was
punctured 5 times for each injection. The test was performed at room
temperature.
|