x pollutant concentration monitor and diluent gas
monitor) will pass the relative accuracy test (see section 6 of this appendix).
It is recommended that monitor measurements be made at locations where the
exhaust gas temperature is above the dew-point temperature. If the cause of
failure to meet the relative accuracy tests is determined to be the measurement
location, relocate the monitor probe(s).
1.1.1 POINT MONITORS
Locate the measurement point (1) within the centroidal area of the stack or
duct cross section, or (2) no less than 1.0 meter from the stack or duct wall.
1.1.2 PATH MONITORS
Locate the measurement path (1) totally within the inner area bounded by a
line 1.0 meter from the stack or duct wall, or (2) such that at least 70.0
percent of the path is within the inner 50.0 percent of the stack or duct
cross-sectional area, or (3) such that the path is centrally located within any
part of the centroidal area.
1.2 Flow Monitors
Install the flow monitor in a location that provides representative
volumetric flow over all operating conditions. Such a location is one that
provides an average velocity of the flue gas flow over the stack or duct cross
section, provides a representative SO2 emission rate (in lb/hr), and
is representative of the pollutant concentration monitor location. Where the
moisture content of the flue gas affects volumetric flow measurements, use the
procedures in both Reference Methods 1 and 4 of appendix A to part 60 of this
chapter to establish a proper location for the flow monitor. The EPA recommends
(but does not require) performing a flow profile study following the procedures
in 40 CFR part 60, appendix A, method, 1, sections 11.5 or 11.4 for each of the
three operating or load levels indicated in section 6.5.2.1 of this appendix to
determine the acceptability of the potential flow monitor location and to
determine the number and location of flow sampling points required to obtain a
representative flow value. The procedure in 40 CFR part 60, appendix A, Test
Method 1, section 11.5 may be used even if the flow measurement location is
greater than or equal to 2 equivalent stack or duct diameters downstream or
greater than or equal to 1/2 duct diameter upstream from a flow disturbance.
If a flow profile study shows that cyclonic (or swirling) or stratified flow
conditions exist at the potential flow monitor location that are likely to
prevent the monitor from meeting the performance specifications of this part,
then EPA recommends either (1) selecting another location where there is no
cyclonic (or swirling) or stratified flow condition, or (2) eliminating the
cyclonic (or swirling) or stratified flow condition by straightening the flow,
e.g., by installing straightening vanes. EPA also recommends selecting flow
monitor locations to minimize the effects of condensation, coating, erosion, or
other conditions that could adversely affect flow monitor performance.
1.2.1 ACCEPTABILITY OF MONITOR
LOCATION
The installation of a flow monitor is acceptable if either (1) the location
satisfies the minimum siting criteria of method 1 in appendix A to part 60 of
this chapter (i.e., the location is greater than or equal to eight stack or duct
diameters downstream and two diameters upstream from a flow disturbance; or, if
necessary, two stack or duct diameters downstream and one-half stack or duct
diameter upstream from a flow disturbance), or (2) the results of a flow profile
study, if performed, are acceptable (i.e., there are no cyclonic (or swirling)
or stratified flow conditions), and the flow monitor also satisfies the
performance specifications of this part. If the flow monitor is installed in a
location that does not satisfy these physical criteria, but nevertheless the
monitor achieves the performance specifications of this part, then the location
is acceptable, notwithstanding the requirements of this section.
1.2.2 ALTERNATIVE MONITORING
LOCATION
Whenever the owner or operator successfully demonstrates that modifications
to the exhaust duct or stack (such as installation of straightening vanes,
modifications of ductwork, and the like) are necessary for the flow monitor to
meet the performance specifications, the Administrator may approve an interim
alternative flow monitoring methodology and an extension to the required
certification date for the flow monitor.
Where no location exists that satisfies the physical siting criteria in
section 1.2.1, where the results of flow profile studies performed at two or
more alternative flow monitor locations are unacceptable, or where installation
of a flow monitor in either the stack or the ducts is demonstrated to be
technically infeasible, the owner or operator may petition the Administrator for
an alternative method for monitoring flow.
2. EQUIPMENT SPECIFICATIONS
2.1 Instrument Span and
Range
In implementing sections 2.1.1 through 2.1.6 of this appendix, set the
measurement range for each parameter (SO2, NOX, CO2, O2, or flow rate) high enough to prevent full-scale
exceedances from occurring, yet low enough to ensure good measurement accuracy
and to maintain a high signal-to-noise ratio. To meet these objectives, select
the range such that the majority of the readings obtained during typical unit
operation are kept, to the extent practicable, between 20.0 and 80.0 percent of
the full-scale range of the instrument. These guidelines do not apply to: (1)
SO2 readings obtained during the combustion of very low sulfur fuel
(as defined in §72.2 of this chapter); (2) SO2 or NOX
readings recorded on the high measurement range, for units with SO2 or
NOX emission controls and two span values, unless the emission
controls are operated seasonally (for example, only during the ozone season); or
(3) SO2 or NOX readings less than 20.0 percent of full-scale
on the low measurement range for a dual span unit, provided that the maximum
expected concentration (MEC), low-scale span value, and low-scale range settings
have been determined according to sections 2.1.1.2, 2.1.1.4(a), (b), and (g) of
this appendix (for SO2), or according to sections 2.1.2.2, 2.1.2.4(a)
and (f) of this appendix (for NOX).
2.1.1 SO2 Pollutant
Concentration Monitors
Determine, as indicated in sections 2.1.1.1 through 2.1.1.5 of this appendix
the span value(s) and range(s) for an SO2 pollutant concentration
monitor so that all potential and expected concentrations can be accurately
measured and recorded. Note that if a unit exclusively combusts fuels that are
very low sulfur fuels (as defined in §72.2 of this chapter), the SO2
monitor span requirements in §75.11(e)(3)(iv) apply in lieu of the requirements
of this section.
2.1.1.1 Maximum Potential
Concentration
(a) Make an initial determination of the maximum potential concentration
(MPC) of SO2 by using Equation A-1a or A-1b. Base the MPC calculation
on the maximum percent sulfur and the minimum gross calorific value (GCV) for
the highest-sulfur fuel to be burned. The maximum sulfur content and minimum GCV
shall be determined from all available fuel sampling and analysis data for that
fuel from the previous 12 months (minimum), excluding clearly anomalous fuel
sampling values. If both the fuel sulfur content and the GCV are routinely
determined from each fuel sample, the owner or operator may, as an alternative
to using the highest individual percent sulfur and lowest individual GCV values
in the MPC calculation, pair the sulfur content and GCV values from each sample
analysis and calculate the ratio of percent sulfur to GCV (i.e., %S/GCV)
for each pair of values. If this option is selected, the MPC shall be calculated
using the highest %S/GCV ratio in Equation A-1a or A-1b. If the designated
representative certifies that the highest-sulfur fuel is never burned alone in
the unit during normal operation but is always blended or co-fired with other
fuel(s), the MPC may be calculated using a best estimate of the highest sulfur
content and lowest gross calorific value expected for the blend or fuel mixture
and inserting these values into Equation A-1a or A-1b. Derive the best estimate
of the highest percent sulfur and lowest GCV for a blend or fuel mixture from
weighted-average values based upon the historical composition of the blend or
mixture in the previous 12 (or more) months. If insufficient representative fuel
sampling data are available to determine the maximum sulfur content and minimum
GCV, use values from contract(s) for the fuel(s) that will be combusted by the
unit in the MPC calculation.
View or
Download PDF
or
View or
Download PDF
Where,
MPC = Maximum potential concentration (ppm, wet basis). (To convert to dry
basis, divide the MPC by 0.9.)
MEC = Maximum expected concentration (ppm, wet basis). (To convert to dry
basis, divide the MEC by 0.9).
%S = Maximum sulfur content of fuel to be fired, wet basis, weight percent,
as determined according to the applicable method in paragraph (c) of section
2.1.1.1.
%O2w = Minimum oxygen concentration, percent wet basis, under
typical operating conditions.
%CO2w = Maximum carbon dioxide concentration, percent wet basis,
under typical operating conditions.
GCV = Minimum gross calorific value of the fuel or blend to be combusted,
based on historical fuel sampling and analysis data or, if applicable, based on
the fuel contract specifications (Btu/lb). If based on fuel sampling and
analysis, the GCV shall be determined according to the applicable method in
paragraph (c) of section 2.1.1.1.
11.32 × 106 = Oxygen-based conversion factor in Btu/lb (ppm)/%.
66.93 × 106 = Carbon dioxide-based conversion factor in Btu/lb
(ppm)/%.
Note: All percent values to be inserted in the equations of this
section are to be expressed as a percentage, not a fractional value (e.g., 3,
not .03).
(b) Alternatively, if a certified SO2 CEMS is already installed,
the owner or operator may make the initial MPC determination based upon
quality-assured historical data recorded by the CEMS. For the purposes of this
section, 2.1.1.1, a "certified" CEMS means a CEM system that has met the
applicable certification requirements of either: This part, or part 60 of this
chapter, or a State CEM program, or the source operating permit. If this option
is chosen, the MPC shall be the maximum SO2 concentration observed
during the previous 720 (or more) quality-assured monitor operating hours when
combusting the highest-sulfur fuel (or highest-sulfur blend if fuels are always
blended or co-fired) that is to be combusted in the unit or units monitored by
the SO2 monitor. For units with SO2 emission controls, the
certified SO2 monitor used to determine the MPC must be located at or
before the control device inlet. Report the MPC and the method of determination
in the monitoring plan required under §75.53. Note that the initial MPC value is
subject to periodic review under section 2.1.1.5 of this appendix. If an MPC
value is found to be either inappropriately high or low, the MPC shall be
adjusted in accordance with section 2.1.1.5, and corresponding span and range
adjustments shall be made, if necessary.
(c) When performing fuel sampling to determine the MPC, use ASTM Methods:
ASTM D3177-89, "Standard Test Methods for Total Sulfur in the Analysis Sample of
Coal and Coke"; ASTM D4239-85, "Standard Test Methods for Sulfur in the Analysis
Sample of Coal and Coke Using High Temperature Tube Furnace Combustion Methods";
ASTM D4294-90, "Standard Test Method for Sulfur in Petroleum Products by
Energy-Dispersive X-Ray Fluorescence Spectroscopy"; ASTM D1552-90, "Standard
Test Method for Sulfur in Petroleum Products (High Temperature Method)"; ASTM
D129-91, "Standard Test Method for Sulfur in Petroleum Products (General Bomb
Method)"; ASTM D2622-92, "Standard Test Method for Sulfur in Petroleum Products
by X-Ray Spectrometry" for sulfur content of solid or liquid fuels; ASTM
D3176-89, "Standard Practice for Ultimate Analysis of Coal and Coke"; ASTM
D240-87 (Reapproved 1991), "Standard Test Method for Heat of Combustion of
Liquid Hydrocarbon Fuels by Bomb Calorimeter"; or ASTM D2015-91, "Standard Test
Method for Gross Calorific Value of Coal and Coke by the Adiabatic Bomb
Calorimeter" for GCV (incorporated by reference under §75.6).
2.1.1.2 Maximum Expected
Concentration
(a) Make an initial determination of the maximum expected concentration (MEC)
of SO2 whenever: (a) SO2 emission controls are used; or (b)
both high-sulfur and low-sulfur fuels (e.g., high-sulfur coal and low-sulfur
coal or different grades of fuel oil) or high-sulfur and low-sulfur fuel blends
are combusted as primary or backup fuels in a unit without SO2
emission controls. For units with SO2 emission controls, use Equation
A-2 to make the initial MEC determination. When high-sulfur and low-sulfur fuels
or blends are burned as primary or backup fuels in a unit without SO2
controls, use Equation A-1a or A-1b to calculate the initial MEC value for each
fuel or blend, except for: (1) the highest-sulfur fuel or blend (for which the
MPC was previously calculated in section 2.1.1.1 of this appendix); (2) fuels or
blends that are very low sulfur fuels (as defined in §72.2 of this chapter); or
(3) fuels or blends that are used only for unit startup. Each initial MEC value
shall be documented in the monitoring plan required under §75.53. Note that each
initial MEC value is subject to periodic review under section 2.1.1.5 of this
appendix. If an MEC value is found to be either inappropriately high or low, the
MEC shall be adjusted in accordance with section 2.1.1.5, and corresponding span
and range adjustments shall be made, if necessary.
(b) For each MEC determination, substitute into Equation A-1a or A-1b the
highest sulfur content and minimum GCV value for that fuel or blend, based upon
all available fuel sampling and analysis results from the previous 12 months (or
more), or, if fuel sampling data are unavailable, based upon fuel contract(s).
(c) Alternatively, if a certified SO2 CEMS is already installed,
the owner or operator may make the initial MEC determination(s) based upon
historical monitoring data. For the purposes of this section, 2.1.1.2, a
"certified" CEMS means a CEM system that has met the applicable certification
requirements of either: This part, or part 60 of this chapter, or a State CEM
program, or the source operating permit. If this option is chosen for a unit
with SO2 emission controls, the MEC shall be the maximum SO2
concentration measured downstream of the control device outlet by the CEMS over
the previous 720 (or more) quality-assured monitor operating hours with the unit
and the control device both operating normally. For units that burn high- and
low-sulfur fuels or blends as primary and backup fuels and have no SO2
emission controls, the MEC for each fuel shall be the maximum SO2
concentration measured by the CEMS over the previous 720 (or more)
quality-assured monitor operating hours in which that fuel or blend was the only
fuel being burned in the unit.
View or
Download PDF
Where:
MEC = Maximum expected concentration (ppm).
MPC = Maximum potential concentration (ppm), as determined by Eq. A-1a or
A-1b in section 2.1.1.1 of this appendix.
RE = Expected average design removal efficiency of control equipment (%).
2.1.1.3 Span Value(s) and
Range(s)
Determine the high span value and the high full-scale range of the SO2 monitor as follows. (Note: For purposes of this part, the high span and
range refer, respectively, either to the span and range of a single span unit or
to the high span and range of a dual span unit.) The high span value shall be
obtained by multiplying the MPC by a factor no less than 1.00 and no greater
than 1.25. Round the span value upward to the next highest multiple of 100 ppm.
If the SO2 span concentration is ≤ 500 ppm, the span value may either
be rounded upward to the next highest multiple of 10 ppm, or to the next highest
multiple of 100 ppm. The high span value shall be used to determine
concentrations of the calibration gases required for daily calibration error
checks and linearity tests. Select the full-scale range of the instrument to be
consistent with section 2.1 of this appendix and to be greater than or equal to
the span value. Report the full-scale range setting and calculations of the MPC
and span in the monitoring plan for the unit. Note that for certain
applications, a second (low) SO2 span and range may be required (see
section 2.1.1.4 of this appendix). If an existing State, local, or federal
requirement for span of an SO2 pollutant concentration monitor
requires or allows the use of a span value lower than that required by this
section or by section 2.1.1.4 of this appendix, the State, local, or federal
span value may be used if a satisfactory explanation is included in the
monitoring plan, unless span and/or range adjustments become necessary in
accordance with section 2.1.1.5 of this appendix. Span values higher than those
required by either this section or section 2.1.1.4 of this appendix must be
approved by the Administrator.
2.1.1.4 Dual Span and Range
Requirements
For most units, the high span value based on the MPC, as determined under
section 2.1.1.3 of this appendix will suffice to measure and record SO2 concentrations (unless span and/or range adjustments become necessary
in accordance with section 2.1.1.5 of this appendix). In some instances,
however, a second (low) span value based on the MEC may be required to ensure
accurate measurement of all possible or expected SO2 concentrations.
To determine whether two SO2 span values are required, proceed as
follows:
(a) For units with SO2 emission controls, compare the MEC from
section 2.1.1.2 of this appendix to the high full-scale range value from section
2.1.1.3 of this appendix. If the MEC is ≥20.0 percent of the high range value,
then the high span value and range determined under section 2.1.1.3 of this
appendix are sufficient. If the MEC is <20.0 percent of the high range value,
then a second (low) span value is required.
(b) For units that combust high- and low-sulfur primary and backup fuels (or
blends) and have no SO2 controls, compare the high range value from
section 2.1.1.3 of this appendix (for the highest-sulfur fuel or blend) to the
MEC value for each of the other fuels or blends, as determined under section
2.1.1.2 of this appendix. If all of the MEC values are ≥20.0 percent of the high
range value, the high span and range determined under section 2.1.1.3 of this
appendix are sufficient, regardless of which fuel or blend is burned in the
unit. If any MEC value is <20.0 percent of the high range value, then a
second (low) span value must be used when that fuel or blend is combusted.
(c) When two SO2 spans are required, the owner or operator may
either use a single SO2 analyzer with a dual range (i.e., low- and
high-scales) or two separate SO2 analyzers connected to a common
sample probe and sample interface. Alternatively, if RATAs are performed and
passed on both measurement ranges, the owner or operator may use two separate
SO2 analyzers connected to separate probes and sample interfaces. For
units with SO2 emission controls, the owner or operator may use a low
range analyzer and a default high range value, as described in paragraph (f) of
this section, in lieu of maintaining and quality assuring a high-scale range.
Other monitor configurations are subject to the approval of the Administrator.
(d) The owner or operator shall designate the monitoring systems and
components in the monitoring plan under §75.53 as follows: when a single probe
and sample interface are used, either designate the low and high monitor ranges
as separate SO2 components of a single, primary SO2
monitoring system; designate the low and high monitor ranges as the SO2 components of two separate, primary SO2 monitoring systems;
designate the normal monitor range as a primary monitoring system and the other
monitor range as a non-redundant backup monitoring system; or, when a single,
dual-range SO2 analyzer is used, designate the low and high ranges as
a single SO2 component of a primary SO2 monitoring system
(if this option is selected, use a special dual-range component type code, as
specified by the Administrator, to satisfy the requirements of
§75.53(e)(1)(iv)(D)). When two SO2 analyzers are connected to separate
probes and sample interfaces, designate the analyzers as the SO2
components of two separate, primary SO2 monitoring systems. For units
with SO2 controls, if the default high range value is used, designate
the low range analyzer as the SO2 component of a primary SO2
monitoring system. Do not designate the default high range as a monitoring
system or component. Other component and system designations are subject to
approval by the Administrator. Note that the component and system designations
for redundant backup monitoring systems shall be the same as for primary
monitoring systems.
(e) Each monitoring system designated as primary or redundant backup shall
meet the initial certification and quality assurance requirements for primary
monitoring systems in §75.20(c) or §75.20(d)(1), as applicable, and appendices A
and B to this part, with one exception: relative accuracy test audits (RATAs)
are required only on the normal range (for units with SO2 emission
controls, the low range is considered normal). Each monitoring system designated
as a non-redundant backup shall meet the applicable quality assurance
requirements in §75.20(d)(2).
(f) For dual span units with SO2 emission controls, the owner or
operator may, as an alternative to maintaining and quality assuring a high
monitor range, use a default high range value. If this option is chosen, the
owner or operator shall report a default SO2 concentration of 200
percent of the MPC for each unit operating hour in which the full-scale of the
low range SO2 analyzer is exceeded.
(g) The high span value and range shall be determined in accordance with
section 2.1.1.3 of this appendix. The low span value shall be obtained by
multiplying the MEC by a factor no less than 1.00 and no greater than 1.25, and
rounding the result upward to the next highest multiple of 10 ppm (or 100 ppm,
as appropriate). For units that burn high- and low-sulfur primary and backup
fuels or blends and have no SO2 emission controls, select, as the
basis for calculating the appropriate low span value and range, the
fuel-specific MEC value closest to 20.0 percent of the high full-scale range
value (from paragraph (b) of this section). The low range must be greater than
or equal to the low span value, and the required calibration gases must be
selected based on the low span value. However, if the default high range option
in paragraph (f) of this section is selected, the full-scale of the low
measurement range shall not exceed five times the MEC value (where the MEC is
rounded upward to the next highest multiple of 10 ppm). For units with two SO2 spans, use the low range whenever the SO2 concentrations are
expected to be consistently below 20.0 percent of the high full-scale range
value, i.e., when the MEC of the fuel or blend being combusted is less than 20.0
percent of the high full-scale range value. When the full-scale of the low range
is exceeded, the high range shall be used to measure and record the SO2 concentrations; or, if applicable, the default high range value in
paragraph (f) of this section shall be reported for each hour of the full-scale
exceedance.
2.1.1.5 Adjustment of Span and
Range
For each affected unit or common stack, the owner or operator shall make a
periodic evaluation of the MPC, MEC, span, and range values for each SO2 monitor (at a minimum, an annual evaluation is required) and shall make
any necessary span and range adjustments, with corresponding monitoring plan
updates, as described in paragraphs (a), (b), and (c) of this section. Span and
range adjustments may be required, for example, as a result of changes in the
fuel supply, changes in the manner of operation of the unit, or installation or
removal of emission controls. In implementing the provisions in paragraphs (a)
and (b) of this section, SO2 data recorded during short-term,
non-representative process operating conditions (e.g., a trial burn of a
different type of fuel) shall be excluded from consideration. The owner or
operator shall keep the results of the most recent span and range evaluation
on-site, in a format suitable for inspection. Make each required span or range
adjustment no later than 45 days after the end of the quarter in which the need
to adjust the span or range is identified, except that up to 90 days after the
end of that quarter may be taken to implement a span adjustment if the
calibration gases currently being used for daily calibration error tests and
linearity checks are unsuitable for use with the new span value.
(a) If the fuel supply, the composition of the fuel blend(s), the emission
controls, or the manner of operation change such that the maximum expected or
potential concentration changes significantly, adjust the span and range setting
to assure the continued accuracy of the monitoring system. A "significant"
change in the MPC or MEC means that the guidelines in section 2.1 of this
appendix can no longer be met, as determined by either a periodic evaluation by
the owner or operator or from the results of an audit by the Administrator. The
owner or operator should evaluate whether any planned changes in operation of
the unit may affect the concentration of emissions being emitted from the unit
or stack and should plan any necessary span and range changes needed to account
for these changes, so that they are made in as timely a manner as practicable to
coordinate with the operational changes. Determine the adjusted span(s) using
the procedures in sections 2.1.1.3 and 2.1.1.4 of this appendix (as applicable).
Select the full-scale range(s) of the instrument to be greater than or equal to
the new span value(s) and to be consistent with the guidelines of section 2.1 of
this appendix.
(b) Whenever a full-scale range is exceeded during a quarter and the
exceedance is not caused by a monitor out-of-control period, proceed as follows:
(1) For exceedances of the high range, report 200.0 percent of the current
full-scale range as the hourly SO2 concentration for each hour of the
full-scale exceedance and make appropriate adjustments to the MPC, span, and
range to prevent future full-scale exceedances.
(2) For units with two SO2 spans and ranges, if the low range is
exceeded, no further action is required, provided that the high range is
available and is not out-of-control or out-of-service for any reason. However,
if the high range is not able to provide quality-assured data at the time of the
low range exceedance or at any time during the continuation of the exceedance,
report the MPC as the SO2 concentration until the readings return to
the low range or until the high range is able to provide quality-assured data
(unless the reason that the high-scale range is not able to provide
quality-assured data is because the high-scale range has been exceeded; if the
high-scale range is exceeded follow the procedures in paragraph (b)(1) of this
section).
(c) Whenever changes are made to the MPC, MEC, full-scale range, or span
value of the SO2 monitor, as described in paragraphs (a) or (b) of
this section, record and report (as applicable) the new full-scale range
setting, the new MPC or MEC and calculations of the adjusted span value in an
updated monitoring plan. The monitoring plan update shall be made in the quarter
in which the changes become effective. In addition, record and report the
adjusted span as part of the records for the daily calibration error test and
linearity check specified by appendix B to this part. Whenever the span value is
adjusted, use calibration gas concentrations that meet the requirements of
section 5.1 of this appendix, based on the adjusted span value. When a span
adjustment is so significant that the calibration gases currently being used for
daily calibration error tests and linearity checks are unsuitable for use with
the new span value, then a diagnostic linearity test using the new calibration
gases must be performed and passed. Use the data validation procedures in
§75.20(b)(3), beginning with the hour in which the span is changed.
2.1.2 NOX Pollutant
Concentration Monitors
Determine, as indicated in sections 2.1.2.1 through 2.1.2.5 of this appendix,
the span and range value(s) for the NOX pollutant concentration
monitor so that all expected NOX concentrations can be determined and
recorded accurately.
2.1.2.1 Maximum Potential
Concentration
(a) The maximum potential concentration (MPC) of NOX for each
affected unit shall be based upon whichever fuel or blend combusted in the unit
produces the highest level of NOX emissions. For the purposes of this
section, 2.1.2.1, and section 2.1.2.2 of this appendix, a "blend" means a
frequently-used fuel mixture having a consistent composition (e.g., an oil and
gas mixture where the relative proportions of the two fuels vary by no more than
10%, on average). Make an initial determination of the MPC using the appropriate
option as follows:
Option 1: Use 800 ppm for coal-fired and 400 ppm for oil- or gas-fired units
as the maximum potential concentration of NOX (if an MPC of 1600 ppm
for coal-fired units or 480 ppm for oil- or gas-fired units was previously
selected under this section, that value may still be used, provided that the
guidelines of section 2.1 of this appendix are met); For cement kilns, use 2000
ppm as the MPC. For process heaters, use 200 ppm if the unit burns only gaseous
fuel and 500 ppm if the unit burns oil;
Option 2: Use the specific values based on boiler type and fuel combusted,
listed in Table 2-1 or Table 2-2; For a new gas-fired or oil-fired combustion
turbine, if a default MPC value of 50 ppm was previously selected from Table
2-2, that value may be used until March 31, 2003;
Option 3: Use NOX emission test results;
Option 4: Use historical CEM data over the previous 720 (or more) unit
operating hours when combusting the fuel or blend with the highest NOX
emission rate; or
Option 5: If a reliable estimate of the uncontrolled NOX emissions
from the unit is available from the manufacturer, the estimated value may be
used.
(b) For the purpose of providing substitute data during NOX missing
data periods in accordance with §§75.31 and 75.33 and as required elsewhere
under this part, the owner or operator shall also calculate the maximum
potential NOX emission rate (MER), in lb/mmBtu, by substituting the
MPC for NOX in conjunction with the minimum expected CO2 or
maximum O2 concentration (under all unit operating conditions except
for unit startup, shutdown, and upsets) and the appropriate F-factor into the
applicable equation in appendix F to this part. The diluent cap value of 5.0
percent CO2 (or 14.0 percent O2) for boilers or 1.0 percent
CO2 (or 19.0 percent O2) for combustion turbines may be used
in the NOX MER calculation. As a second alternative, when the NOX MPC is determined from emission test results or from historical CEM
data, as described in paragraphs (a), (d) and (e) of this section,
quality-assured diluent gas (i.e., O2 or CO2) data recorded
concurrently with the MPC may be used to calculate the MER.
(c) Report the method of determining the initial MPC and the calculation of
the maximum potential NOX emission rate in the monitoring plan for the
unit. Note that whichever MPC option in paragraph 2.1.2.1(a) of this appendix is
selected, the initial MPC value is subject to periodic review under section
2.1.2.5 of this appendix. If an MPC value is found to be either inappropriately
high or low, the MPC shall be adjusted in accordance with section 2.1.2.5, and
corresponding span and range adjustments shall be made, if necessary.
(d) For units with add-on NOX controls (whether or not the unit is
equipped with low-NOX burner technology), or for units equipped with
dry low-NOX (DLN) technology, NOX emission testing may only
be used to determine the MPC if testing can be performed either upstream of the
add-on controls or during a time or season when the add-on controls are not in
operation or when the DLN controls are not in the premixed (low-NOX)
mode. If NOX emission testing is performed, use the following
guidelines. Use Method 7E from appendix A to part 60 of this chapter to measure
total NOX concentration. (Note: Method 20 from appendix A to part 60
may be used for gas turbines, instead of Method 7E.) Operate the unit, or group
of units sharing a common stack, at the minimum safe and stable load, the normal
load, and the maximum load. If the normal load and maximum load are identical,
an intermediate level need not be tested. Operate at the highest excess O2 level expected under normal operating conditions. Make at least three
runs of 20 minutes (minimum) duration with three traverse points per run at each
operating condition. Select the highest point NOX concentration from
all test runs as the MPC for NOX.
(e) If historical CEM data are used to determine the MPC, the data must, for
uncontrolled units or units equipped with low-NOX burner technology
and no other NOX controls, represent a minimum of 720 quality-assured
monitor operating hours from the NOX component of a certified
monitoring system, obtained under various operating conditions including the
minimum safe and stable load, normal load (including periods of high excess air
at normal load), and maximum load. For the purposes of this section, 2.1.2.1, a
"certified" CEMS means a CEM system that has met the applicable certification
requirements of either: this part, or part 60 of this chapter, or a State CEM
program, or the source operating permit. For a unit with add-on NOX
controls (whether or not the unit is equipped with low-NOX burner
technology), or for a unit equipped with dry low-NOX (DLN) technology,
historical CEM data may only be used to determine the MPC if the 720
quality-assured monitor operating hours of CEM data are collected upstream of
the add-on controls or if the 720 hours of data include periods when the add-on
controls are not in operation or when the DLN controls are not in the premixed
(low-NOX mode). For units that do not produce electrical or thermal
output, the data must represent the full range of normal process operation. The
highest hourly NOX concentration in ppm shall be the MPC. Table 2-1_Maximum Potential Concentration for NO[INF]X[/INF]_Coal-Fired
Units
------------------------------------------------------------------------
Maximum
potential
Unit type concentration
for NO[INF]X[/
INF] (ppm)
------------------------------------------------------------------------
Tangentially-fired dry bottom and fluidized bed......... 460
Wall-fired dry bottom, turbo-fired dry bottom, stokers.. 675
Roof-fired (vertically-fired) dry bottom, cell burners, 975
arch-fired.............................................
Cyclone, wall-fired wet bottom, wet bottom turbo-fired.. 1200
Others.................................................. (\1\)
------------------------------------------------------------------------
\1\ As approved by the Administrator.
View or
Download PDF
2.1.2.2 Maximum Expected
Concentration
(a) Make an initial determination of the maximum expected concentration (MEC)
of NOX during normal operation for affected units with add-on NOX controls of any kind (e.g., steam injection, water injection,
SCR, or SNCR) and for turbines that use dry low-NOX technology.
Determine a separate MEC value for each type of fuel (or blend) combusted in the
unit, except for fuels that are only used for unit startup and/or flame
stabilization. Calculate the MEC of NOX using Equation A-2, if
applicable, inserting the maximum potential concentration, as determined using
the procedures in section 2.1.2.1 of this appendix. Where Equation A-2 is not
applicable, set the MEC either by: (1) measuring the NOX concentration
using the testing procedures in this section; (2) using historical CEM data over
the previous 720 (or more) quality-assured monitor operating hours; or (3) if
the unit has add-on NOX controls or uses dry low NOX
technology, and has a federally-enforceable permit limit for NOX
concentration, the permit limit may be used as the MEC. Include in the
monitoring plan for the unit each MEC value and the method by which the MEC was
determined. Note that each initial MEC value is subject to periodic review under
section 2.1.2.5 of this appendix. If an MEC value is found to be either
inappropriately high or low, the MEC shall be adjusted in accordance with
section 2.1.2.5, and corresponding span and range adjustments shall be made, if
necessary.
(b) If NOX emission testing is used to determine the MEC value(s),
the MEC for each type of fuel (or blend) shall be based upon testing at minimum
load, normal load, and maximum load. At least three tests of 20 minutes
(minimum) duration, using at least three traverse points, shall be performed at
each load, using Method 7E from appendix A to part 60 of this chapter (Note:
Method 20 from appendix A to part 60 may be used for gas turbines instead of
Method 7E). The test must be performed at a time when all NOX control
devices and methods used to reduce NOX emissions (if applicable) are
operating properly. The testing shall be conducted downstream of all NOX controls. The highest point NOX concentration (e.g., the
highest one-minute average) recorded during any of the test runs shall be the
MEC.
(c)If historical CEM data are used to determine the MEC value(s), the MEC for
each type of fuel shall be based upon 720 (or more) hours of quality-assured
data from the NOX component of a certified monitoring system
representing the entire load range under stable operating conditions. For the
purposes of this section, 2.1.2.2, a "certified" CEMS means a CEM system that
has met the applicable certification requirements of either: this part, or part
60 of this chapter, or a State CEM program, or the source operating permit. The
data base for the MEC shall not include any CEM data recorded during unit
startup, shutdown, or malfunction or (for units with add-on NOX
controls or turbines using dry low NOX technology) during any NOX control device malfunctions or outages. All NOX control
devices and methods used to reduce NOX emissions (if applicable) must
be operating properly during each hour. The CEM data shall be collected
downstream of all NOX controls. For each type of fuel, the highest of
the 720 (or more) quality-assured hourly average NOX concentrations
recorded by the CEMS shall be the MEC.
2.1.2.3 Span Value(s) and
Range(s)
(a) Determine the high span value of the NOX monitor as follows.
The high span value shall be obtained by multiplying the MPC by a factor no less
than 1.00 and no greater than 1.25. Round the span value upward to the next
highest multiple of 100 ppm. If the NOX span concentration is ≤500
ppm, the span value may either be rounded upward to the next highest multiple of
10 ppm, or to the next highest multiple of 100 ppm. The high span value shall be
used to determine the concentrations of the calibration gases required for daily
calibration error checks and linearity tests. Note that for certain
applications, a second (low) NOX span and range may be required (see
section 2.1.2.4 of this appendix).
(b) If an existing State, local, or federal requirement for span of a NOX pollutant concentration monitor requires or allows the use of a span
value lower than that required by this section or by section 2.1.2.4 of this
appendix, the State, local, or federal span value may be used, where a
satisfactory explanation is included in the monitoring plan, unless span and/or
range adjustments become necessary in accordance with section 2.1.2.5 of this
appendix. Span values higher than required by this section or by section 2.1.2.4
of this appendix must be approved by the Administrator.
(c) Select the full-scale range of the instrument to be consistent with
section 2.1 of this appendix and to be greater than or equal to the high span
value. Include the full-scale range setting and calculations of the MPC and span
in the monitoring plan for the unit.
2.1.2.4 Dual Span and Range
Requirements
For most units, the high span value based on the MPC, as determined under
section 2.1.2.3 of this appendix will suffice to measure and record NOX concentrations (unless span and/or range adjustments must be made in
accordance with section 2.1.2.5 of this appendix). In some instances, however, a
second (low) span value based on the MEC may be required to ensure accurate
measurement of all expected and potential NOX concentrations. To
determine whether two NOX spans are required, proceed as follows:
(a) Compare the MEC value(s) determined in section 2.1.2.2 of this appendix
to the high full-scale range value determined in section 2.1.2.3 of this
appendix. If the MEC values for all fuels (or blends) are ≥20.0 percent of the
high range value, the high span and range values determined under section
2.1.2.3 of this appendix are sufficient, irrespective of which fuel or blend is
combusted in the unit. If any of the MEC values is <20.0 percent of the high
range value, two spans (low and high) are required, one based on the MPC and the
other based on the MEC.
(b) When two NOX spans are required, the owner or operator may
either use a single NOX analyzer with a dual range (low-and
high-scales) or two separate NOX analyzers connected to a common
sample probe and sample interface. Two separate NOX analyzers
connected to separate probes and sample interfaces may be used if RATAs are
passed on both ranges. For units with add-on NOX emission controls
(e.g., steam injection, water injection, SCR, or SNCR) or units equipped with
dry low-NOX technology, the owner or operator may use a low range
analyzer and a "default high range value," as described in paragraph 2.1.2.4(e)
of this section, in lieu of maintaining and quality assuring a high-scale range.
Other monitor configurations are subject to the approval of the Administrator.
(c) The owner or operator shall designate the monitoring systems and
components in the monitoring plan under §75.53 as follows: when a single probe
and sample interface are used, either designate the low and high ranges as
separate NOX components of a single, primary NOX monitoring
system; designate the low and high ranges as the NOX components of two
separate, primary NOX monitoring systems; designate the normal range
as a primary monitoring system and the other range as a non-redundant backup
monitoring system; or, when a single, dual-range NOX analyzer is used,
designate the low and high ranges as a single NOX component of a
primary NOX monitoring system (if this option is selected, use a
special dual-range component type code, as specified by the Administrator, to
satisfy the requirements of §75.53(e)(1)(iv)(D)). When two NOX
analyzers are connected to separate probes and sample interfaces, designate the
analyzers as the NOX components of two separate, primary NOX
monitoring systems. For units with add-on NOX controls or units
equipped with dry low-NOX technology, if the default high range value
is used, designate the low range analyzer as the NOX component of the
primary NOX monitoring system. Do not designate the default high range
as a monitoring system or component. Other component and system designations are
subject to approval by the Administrator. Note that the component and system
designations for redundant backup monitoring systems shall be the same as for
primary monitoring systems.
(d) Each monitoring system designated as primary or redundant backup shall
meet the initial certification and quality assurance requirements in §75.20(c)
(for primary monitoring systems), in §75.20(d)(1) (for redundant backup
monitoring systems) and appendices A and B to this part, with one exception:
relative accuracy test audits (RATAs) are required only on the normal range (for
dual span units with add-on NOX emission controls, the low range is
considered normal). Each monitoring system designated as non-redundant backup
shall meet the applicable quality assurance requirements in §75.20(d)(2).
(e) For dual span units with add-on NOX emission controls (e.g.,
steam injection, water injection, SCR, or SNCR), or, for units that use dry low
NOX technology, the owner or operator may, as an alternative to
maintaining and quality assuring a high monitor range, use a default high range
value. If this option is chosen, the owner or operator shall report a default
value of 200.0 percent of the MPC for each unit operating hour in which the
full-scale of the low range NOX analyzer is exceeded.
(f) The high span and range shall be determined in accordance with section
2.1.2.3 of this appendix. The low span value shall be 100.0 to 125.0 percent of
the MEC, rounded up to the next highest multiple of 10 ppm (or 100 ppm, if
appropriate). If more than one MEC value (as determined in section 2.1.2.2 of
this appendix) is <20.0 percent of the high full-scale range value, the low
span value shall be based upon whichever MEC value is closest to 20.0 percent of
the high range value. The low range must be greater than or equal to the low
span value, and the required calibration gases for the low range must be
selected based on the low span value. However, if the default high range option
in paragraph (e) of this section is selected, the full-scale of the low
measurement range shall not exceed five times the MEC value (where the MEC is
rounded upward to the next highest multiple of 10 ppm). For units with two NOX spans, use the low range whenever NOX concentrations are
expected to be consistently <20.0 percent of the high range value, i.e., when
the MEC of the fuel being combusted is <20.0 percent of the high range value.
When the full-scale of the low range is exceeded, the high range shall be used
to measure and record the NOX concentrations; or, if applicable, the
default high range value in paragraph (e) of this section shall be reported for
each hour of the full-scale exceedance.
2.1.2.5 Adjustment of Span and
Range
For each affected unit or common stack, the owner or operator shall make a
periodic evaluation of the MPC, MEC, span, and range values for each NOX monitor (at a minimum, an annual evaluation is required) and shall make
any necessary span and range adjustments, with corresponding monitoring plan
updates, as described in paragraphs (a), (b), and (c) of this section. Span and
range adjustments may be required, for example, as a result of changes in the
fuel supply, changes in the manner of operation of the unit, or installation or
removal of emission controls. In implementing the provisions in paragraphs (a)
and (b) of this section, note that NOX data recorded during
short-term, non-representative operating conditions (e.g., a trial burn of a
different type of fuel) shall be excluded from consideration. The owner or
operator shall keep the results of the most recent span and range evaluation
on-site, in a format suitable for inspection. Make each required span or range
adjustment no later than 45 days after the end of the quarter in which the need
to adjust the span or range is identified, except that up to 90 days after the
end of that quarter may be taken to implement a span adjustment if the
calibration gases currently being used for daily calibration error tests and
linearity checks are unsuitable for use with the new span value.
(a) If the fuel supply, emission controls, or other process parameters change
such that the maximum expected concentration or the maximum potential
concentration changes significantly, adjust the NOX pollutant
concentration span(s) and (if necessary) monitor range(s) to assure the
continued accuracy of the monitoring system. A "significant" change in the MPC
or MEC means that the guidelines in section 2.1 of this appendix can no longer
be met, as determined by either a periodic evaluation by the owner or operator
or from the results of an audit by the Administrator. The owner or operator
should evaluate whether any planned changes in operation of the unit or stack
may affect the concentration of emissions being emitted from the unit and should
plan any necessary span and range changes needed to account for these changes,
so that they are made in as timely a manner as practicable to coordinate with
the operational changes. An example of a change that may require a span and
range adjustment is the installation of low-NOX burner technology on a
previously uncontrolled unit. Determine the adjusted span(s) using the
procedures in section 2.1.2.3 or 2.1.2.4 of this appendix (as applicable).
Select the full-scale range(s) of the instrument to be greater than or equal to
the adjusted span value(s) and to be consistent with the guidelines of section
2.1 of this appendix.
(b) Whenever a full-scale range is exceeded during a quarter and the
exceedance is not caused by a monitor out-of-control period, proceed as follows:
(1) For exceedances of the high range, report 200.0 percent of the current
full-scale range as the hourly NOX concentration for each hour of the
full-scale exceedance and make appropriate adjustments to the MPC, span, and
range to prevent future full-scale exceedances.
(2) For units with two NOX spans and ranges, if the low range is
exceeded, no further action is required, provided that the high range is
available and is not out-of-control or out-of-service for any reason. However,
if the high range is not able to provide quality-assured data at the time of the
low range exceedance or at any time during the continuation of the exceedance,
report the MPC as the NOX concentration until the readings return to
the low range or until the high range is able to provide quality-assured data
(unless the reason that the high-scale range is not able to provide
quality-assured data is because the high-scale range has been exceeded; if the
high-scale range is exceeded, follow the procedures in paragraph (b)(1) of this
section).
(c) Whenever changes are made to the MPC, MEC, full-scale range, or span
value of the NOX monitor as described in paragraphs (a) and (b) of
this section, record and report (as applicable) the new full-scale range
setting, the new MPC or MEC, maximum potential NOX emission rate, and
the adjusted span value in an updated monitoring plan for the unit. The
monitoring plan update shall be made in the quarter in which the changes become
effective. In addition, record and report the adjusted span as part of the
records for the daily calibration error test and linearity check required by
appendix B to this part. Whenever the span value is adjusted, use calibration
gas concentrations that meet the requirements of section 5.1 of this appendix,
based on the adjusted span value. When a span adjustment is significant enough
that the calibration gases currently being used for daily calibration error
tests and linearity checks are unsuitable for use with the new span value, a
diagnostic linearity test using the new calibration gases must be performed and
passed. Use the data validation procedures in §75.20(b)(3), beginning with the
hour in which the span is changed.
2.1.3 CO2 and O2 Monitors
* * * If a dual-range or
autoranging diluent analyzer is installed, the analyzer may be represented in
the monitoring plan as a single component, using a special component type code
specified by the Administrator to satisfy the requirements of
§75.53(e)(1)(iv)(D).
2.1.3 CO2 and O2 Monitors
For an O2 monitor (including O2 monitors used to measure
CO2 emissions or percentage moisture), select a span value between
15.0 and 25.0 percent O2. For a CO2 monitor installed on a
boiler, select a span value between 14.0 and 20.0 percent CO2. For a
CO2 monitor installed on a combustion turbine, an alternative span
value between 6.0 and 14.0 percent CO2 may be used. An alternative O